3-16 Marker Menu
Interference Analyzer (Option 25)
3-72
PN: 10580-00349 Rev. M
Spectrum Analyzer MG
Marker and Peak Menu
Key Sequence:
Marker >
More Peak Options
Peak Search:
Places the currently active marker on the highest amplitude
signal currently on screen.
Next Peak Left:
From the current position of the active marker, the
instrument searches to the left (toward lower frequencies) for a peak signal
that rises at least a certain amount above the average noise level. If no such
peak is found, the marker is placed at the left end of the trace. The Peak
Threshold key allows the user to specify the performance of peak searching.
Next Peak Right:
From the current position of the active marker, the
instrument searches to the right (toward higher frequencies) for a peak signal
that rises at least a certain amount above the average noise level. If no such
peak is found, the marker is placed at the right end of the trace. The Peak
Threshold key allows the user to specify the performance of peak searching.
Delta Marker to Span:
Sets the total span width to the value of the delta
marker. If the delta marker is zero, the span is set to 10 Hz. If there is no delta
marker, or the delta marker value is set to less than 10 Hz, then the span will
be set to 10 Hz.
Marker Freq to Center:
Sets the center frequency to the frequency of the
currently active marker.
Marker to Ref Lvl:
Sets the reference level, top graticule line, to the
amplitude of the currently active marker.
Peak Threshold:
Sets the percentage of the current display that a signal
must rise before it is considered a peak. For example, 10% is one division of
the ten division graticule.
Back:
Returns to the
Figure 3-66.
IA Marker & Peak Menu
Marker & Peak
Back
Next Peak
Left
Next Peak
Right
Delta Marker
to
Span
Marker Freq
to
Center
Marker
to
Ref Lvl
Peak Threshold
10.00%
Peak Search
Содержание BTS Master
Страница 103: ...Spectrum Analyzer 2 35 Trace Menu Spectrum Analyzer MG PN 10580 00349 Rev M 2 91 Figure 2 78 Trace Info...
Страница 246: ...6 6 Coverage Mapping Menu Coverage Mapping Option 431 6 24 PN 10580 00349 Rev M Spectrum Analyzer MG...
Страница 278: ...7 21 Marker Menu AM FM PM Analyzer Option 509 7 32 PN 10580 00349 Rev M Spectrum Analyzer MG...
Страница 282: ...8 2 Bias Tee Menu Bias Tee Option 10 8 4 PN 10580 00349 Rev M Spectrum Analyzer MG...