6.9 Measuring
Waveform
6-119
6
How to Operate Sampling Scope
Note:
Up to 1023 characters can be entered. However, all characters may
not be displayed because the characters that can be displayed are
limited.
The color of the label cannot be changed.
Deleting the displayed label
The channels subject to label deletion vary depending on the Control Ch
setting.
All
Deletes labels for all channels.
Single
Deletes the label for the active channel.
1. Click
Setup to open the Setup dialog box.
2. Click
Utilities tab.
3. Click
Delete for Label, and the displayed label is deleted.
6.9.9
Saving Measurement Results
The sampling oscilloscope results can be saved in the following file types:
●
Screen image
The image of the entire screen or only the measurement screen is
saved to a PNG or JPEG file.
●
Measurement result
The waveform is saved as a text file or CSV file.
●
Measurement conditions
The setting values in the Amplitude O/E, Measurement, Setup, and
Time CRU dialog boxes are saved to a file.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...