Chapter 3 Measurement Examples
3-4
3.2 Measuring Waveform
When measuring the DUT with electrical signal inputs and outputs:
Input the output signal of the built-in PPG to the DUT, and then measure
the DUT output using the sampling oscilloscope.
Figure 3.2-1 When Measuring DUT With Electrical Signal Inputs and
Outputs (MP2110A-012, 021)
1. Connect the DUT input connectors to PPG1
Data
Out
and
Data
Out
using coaxial cables.
If the DUT has only one input connector, connect it to the PPG1
Data
Out
. Also, connect the coaxial terminator included as a standard
accessory to the
Data
Out
.
2. Connect the
Clk Out
(or
Sync Out
) on BERT to
Trigger Clk In
on
Scope using a coaxial cable.
When the
Sync Out
is connected, connect the supplied coaxial
terminator to
Sync
Out
.
3. Connect the DUT output connectors to
Ch A In
and
Ch B In
on the
sampling oscilloscope using coaxial cables.
When the amplitude of the input signal is greater than 400 mVp-p,
connect the attenuators to
Ch A In
and
Ch B In
for more accurate
measurement.
DUT
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...