7.8 Setting Pattern Length
7-35
7
Measurin
g W
aveform
7.8 Setting Pattern Length
When the Pulse mode is set as described in Section 7.7 “Setting Data
Collection Method”, the pattern length can be set.
In the Pulse mode, the pattern is synchronized by collecting data at the
pattern length time cycle.
1. Touch
[Time].
2. Touch the Tracking button at Pattern Length to set to [Off].
3. Touch the Length text box at Pattern Length.
4. Input the pattern length by the bit unit.
Note:
The settable pattern length is up to 1677216 (2
24
).
When Text Pattern at PPG/ED is 2^31–1, [Tracking On] cannot be
set to Tracking at Pattern Length.
When Pattern Length is set to [Tracking On], Test Pattern at
PPG/ED cannot be set to 2^31–1.
When changing the PPG/ED pattern length settings using the
BERTWave, the changed setting values can be set to the sampling
oscilloscope automatically.
1. Touch
[Time].
2. Touch the Tracking button at Pattern Length, and then set to [On].
3. Touch the Master button at Pattern Length.
4. Select the PPG and ED with equal pattern lengths.
Содержание BERTWave MP2100B
Страница 24: ...xxiv...
Страница 96: ...Chapter 2 Before Use 2 42...
Страница 112: ...Chapter 3 Connecting with DUT 3 16...
Страница 250: ...Chapter 7 Measuring Waveform 7 64 Figure 7 12 1 Marker Display...
Страница 262: ...Chapter 7 Measuring Waveform 7 76...
Страница 276: ...Chapter 8 Operating Optical Interface 8 14...
Страница 304: ...Chapter 9 Performance Test 9 28...
Страница 313: ...10 4 Displaying Software Version 10 9 10 Maintenance 4 Touch Exit and then the software version display screen is closed...
Страница 320: ...Chapter 10 Maintenance 10 16...
Страница 374: ...Appendix E Performance Test Record Form E 10...
Страница 382: ...Index Index 6...