7.5 Calibrating
Sampling
Oscilloscope
7-21
7
Measurin
g W
aveform
7.5 Calibrating Sampling Oscilloscope
7.5.1 Calibrating Level
The amplitude accuracy of the sampling oscilloscope is guaranteed after
performing calibration.
When the calibration is required, the error message "Calibration is
required" is displayed in the Setup dialog box in red.
Perform the calibration, when using the BERTWave for the first time or
the message is displayed.
The calibration procedures are as follows:
1. Check that no signals are input to the front-panel Ch A In, Ch B In,
or Trigger Clk In connectors.
2. Touch [Setup] to open the Setup dialog box.
3. Touch the [Utility] tab.
4. Touch [Calibration], and then the message is displayed to confirm
that the signal is not input to the connector.
5. Touch
[OK].