1-12
RECOMMENDED TEST
EQUIPMENT
Table 1-2 provides a list of recommended test equipment needed for
the performance verification, calibration, and troubleshooting proce-
dures presented in this manual.
1-12
690XXB/691XXB MM
INSTRUMENT
CRITICAL
SPECIFICATION
RECOMMENDED
MANUFACTURER/MODEL
USAGE
(1)
Spectrum Analyzer,
with
External Mixers
and
Diplexer Assy
Frequency Range: 0.01 to 65 GHz
Resolution Bandwidth: 10 Hz
Tektronix, Model 2794, with
External Mixers:
WM780K (18 to 26.5 GHz)
WM780A (26.5 to 40 GHz)
WM780U (40 to 60 GHz)
WM780E (60 to 90 GHz)
Diplexer Assy: 015-0385-00
C, P, T
Phase Noise
Measurement
System
Frequency Range: 5 MHz to 26.5 GHz
Aeroflex/Comstron PN9000
with
PN9060-00 Status Module
PN9470-00 Noise Output Module
PN9450-00 Lock Control Module
PN9342-01 Phase Detector Module
PN9530-00 Crystal Oscillator Module
P
Frequency Counter
Frequency Range: 0.01 to 40 GHz
Input Impedance: 50
W
Resolution: 1 Hz
Other: External Time Base Input
Anritsu Model MF2414A
C, P
Power Meter,
with
Power
Sensors
Power Range: –30 to +20 dBm
(1
m
W to 100mW)
Anritsu Model ML2437A or ML2438A,
with
Power Sensors:
MA2474A (0.01 to 40 GHz)
MA2475A (0.01 to 50 GHz)
C, P
Digital Multimeter
Resolution: 4-1/2 digits (to 20V)
DC Accuracy: 0.002% +2 counts
DC Input Impedance: 10 M
W
AC Accuracy: 0.07% +100 counts
(to 20 kHz)
AC Input Impedance: 1 M
W
John Fluke, Inc., Model 8840A, with
Option 8840A-09K (True RMS AC)
C, T
Frequency Reference
Frequency: 10 MHz
Accuracy: 5 x 10
-12
parts/day
Absolute Time Corp., Model 300
P
Function Generator
Output Voltage: 2 volts peak-to-peak
Functions: 0.4 Hz to 100 kHz sine and
square waveforms
Hewlett-Packard, Model 33120A
C, P
Modulation Analyzer
Frequency Input: 10 MHz
(of the IF of the spectrum analyzer)
AM Depth: 0% to 90%
AM Modulation Rates: DC to 100 kHz
Filters: 20 kHz lowpass, 300 Hz highpass
Hewlett-Packard, Model 8901A
P
Table 1-2.
Recommended Test Equipment (1 of 2)
GENERAL
RECOMMENDED
INFORMATION
TEST EQUIPMENT
Содержание 680 C Series
Страница 4: ......
Страница 5: ......
Страница 13: ...Figure 1 1 Typical Series 690XXB 691XXB Synthesized CW Signal Generator Model 69187B Shown ...
Страница 61: ......
Страница 97: ......
Страница 205: ......
Страница 207: ......
Страница 221: ......
Страница 225: ......
Страница 241: ......
Страница 259: ......
Страница 275: ......
Страница 285: ......
Страница 289: ......
Страница 299: ......
Страница 303: ......
Страница 315: ......