8-4 W-CDMA EMF Menu
EMF (Option 444)
8-10
PN: 10580-00234 Rev. Y
3GPP MG
8-4
W-CDMA EMF Menu
Key Sequence:
Measurements
> OTA > EMF
Measurement On/Off:
Starts the EMF Measurement and removes access to
all other menu buttons. The measurement turns On only if the Center
Frequency is set within the valid range and the Anritsu Isotropic Antenna is
connected.
Note that the Measurement Time and other related parameters must be set
before starting the measurement. This button is useful for stopping or
restarting measurements when settings need to be changed. When the
measurement is in progress, access to other menus and key presses are
blocked.
Measurement Time:
Sets the duration of each EMF measurement from one
minute up to 30 minutes. The default is 6 min. The instrument captures
over-the-air data for the X axis when a sync signal is found and there is a valid
scrambling code, then moves to the Y and Z axes. There is no axis dwell time
parameter. You will get as many isotropic results for the set of three axes as
can be obtained within the specified Measurement Time.
When no valid sync signal is found for the current axis, data captured for this
axis will be excluded from the measurement results and the instrument moves
to the next axis. Refer to
“Measurement Results (W-CDMA)” on page 8-14
Number of Measurements:
Sets the number of EMF measurements to
complete from 1 up to 10,000. The EMF test is fully executed when the
specified number of measurements have completed.
Auto Logging On/Off:
Auto Logging is On by default. This must be selected
prior to starting the measurements for the results to be logged. The PCPICH
(Primary Common Pilot Channel) and Field Strength actual and average
values of the current measurement are stored. The maximum, minimum, and
total average values stored are not per measurement, but for all the values
captured until that point.
The location of the saved log file is a new folder named with the current time
stamp followed by _1, and created in “/Internal Memory/EMF/”. The folder can
hold 100 files. Each file holds five measurements. The 101
st
file and the files
created thereafter are stored in a new folder with the same time stamp as the
first, followed by _2 (then _3, and so on). Each file has its own time stamp.
Measurement Parameters:
Opens the
.
Figure 8-9.
W-CDMA EMF Menu (1 of 2)
EMF
Measurement
On Off
EMF Units
V/m
Number of
Measurements
5
Measurement Time
6 min
Auto Logging
On Off
Limits
6.00 V/m
Measurement
Parameters
Back
Содержание 3GPP
Страница 32: ...2 18 Other Menus GSM GPRS EDGE Signal Analyzer 2 18 PN 10580 00234 Rev Y 3GPP MG...
Страница 76: ...3 18 Other Menus W CDMA HSPA Signal Analyzer 3 44 PN 10580 00234 Rev Y 3GPP MG...
Страница 104: ...4 16 Other Menus TD SCDMA HSPA Signal Analyzer 4 28 PN 10580 00234 Rev Y 3GPP MG...
Страница 152: ...5 19 Other Menus LTE Signal Analyzer 5 48 PN 10580 00234 Rev Y 3GPP MG...
Страница 204: ...6 19 Other Menus TD LTE Signal Analyzer 6 52 PN 10580 00234 Rev Y 3GPP MG...
Страница 220: ...7 5 Coverage Mapping Menu LTE and TD LTE Coverage Mapping 7 16 PN 10580 00234 Rev Y 3GPP MG...
Страница 236: ...8 4 W CDMA EMF Menu EMF Option 444 8 16 PN 10580 00234 Rev Y 3GPP MG...
Страница 242: ...Index 4 PN 10580 00234 Rev Y 3GPP MG...
Страница 243: ......