LRL/LRM Calibration
(Waveguide)
The waveguide procedure is very similar to the coaxial and microstrip
procedures already described.
Step 1.
Same as Steps 1 through 6 in the Microstrip
procedure, except choose WAVEGUIDE in menu
C11A.
The only difference is with menu C3F (left). For a
waveguide calibration, move the cursor to WAVE-
GUIDE CUTOFF FREQ and press
Enter
. This ac-
tion calls menu C15B, which lets you enter the
waveguide cutoff frequency. After doing, so you are
returned to menu C3F.
Step 2.
When menu C3F reappears, place cursor on
CHANGE LRL/LRM PARAMETERS and press
the
Enter
key.
Step 3.
Same as Steps 9 through 13 in the Microstrip
procedure.
LRL/LRM CALIBRATION
MEASUREMENT CALIBRATION
7-42
37XXXC OM
MENU C3F
CONFIRM
CALIBRATION
PARAMETERS
LRL/LRM
PARAMETERS
WAVEGUIDE
CUTOFF FREQ
TEST SIGNALS
START CAL
PRESS <ENTER>
TO SELECT
OR CHANGE
MENU C15B
ENTER
WAVEGUIDE
CUTOFF
FREQUENCY
WAVEGUIDE
CUTOFF FREQ
XX.XXXX GHz
PRESS <ENTER>
WHEN COMPLETE
Содержание 37 C Series
Страница 3: ......
Страница 4: ......
Страница 18: ...Figure 1 1 Model 37XXXC Vector Network Analyzer System...
Страница 32: ......
Страница 41: ......
Страница 61: ...Intentionally Blank...
Страница 96: ...Table of Contents 5 1 INTRODUCTION 5 3 5 2 ERROR MESSAGES 5 3 Chapter 5 Error and Status Messages...
Страница 97: ......
Страница 103: ......
Страница 117: ...HARD COPY AND DISK OUTPUT DATA DISPLAYS 6 16 37XXXC OM Figure 6 12 Example of a Tabular Printout 37347C...
Страница 223: ......
Страница 242: ...37XXXC OM 9 21 9 22 TIME DOMAIN MENU FLOWCHART Figure 9 25 Time Domain Menu Flow...
Страница 276: ......
Страница 323: ......
Страница 567: ......
Страница 582: ...37100C 37200C 37300C Vector Network Analyzers Technical Data Sheet Vector Network Analysis up to 65 GHz...