3-11 Matched Thru Tests – Port 1/Port 2
Using the 2300-580-R Software with VectorStar ME7838D4
3-30
PN: 10410-00327 Rev. D
3659 0.8 mm Cal/Ver Kit & 2300-580-R PVS UG
3-11 Matched Thru Tests – Port 1/Port 2
If the
Run All Tests
button on the
Application Interface - Main Menu
tab was selected, the
Matched Thru
tests (in
this section) are executed, followed by the
Mismatched Thru
tests described in
Tests – Port 1/Port 2” on page 3-33
If the
Matched Thru
test button was selected, only the steps in this section are executed. When this test is
complete, the program returns to the main menu described
Section 3-9 “Application Interface – Main Menu
Procedure
1.
The
Enter Serial Number Match Thru
dialog box appears.
2.
Ensure that the correct device is connected.
•
The
Matched Thru
verification standard is the device with one scribe line.
•
The
Mismatched Thru
verification standard is the device with two scribe lines.
3.
Enter the serial number of the Matched Thru.
4.
When ready to proceed, click
OK
. The
Connect Mismatched Thru Male Connector
dialog box appears.
5.
Connect the Matched Thru male connector to VNA Port 1 with the label facing up.
Note
The verification devices must be connected to the
0.8 mm Connector Coupler Ports
and not to the
VNA Ports
. Use the torque wrench supplied with the verification kit when tightening the device
connections.
Figure 3-43.
Enter Serial Number Thru Dialog
Note
The verification standards must be connected to the
0.8 mm Coupler Ports
and not to the
VNA
ports
.
Figure 3-44.
Connect the Mismatched Thru Dialog