Using the 2300-580-R Software with VectorStar ME7838D
2-11 Matched Thru Tests
3659 0.8 mm Cal/Ver Kit & 2300-580-R PVS UG
PN: 10410-00327 Rev. D
2-29
2-11 Matched Thru Tests
If the
Run All Tests
button on the
Application Interface - Main Menu
tab was selected, the
Matched Thru
tests
(immediately below) are executed followed by the
Mismatched Thru
tests described in
If the
Matched Thru
test button was selected, only the steps in this section are executed. When this test is
complete, the program returns to the main menu described
Section 2-9 “Application Interface – Main Menu
above.
Procedure
1.
The
Enter Serial Number Match Thru
dialog box appears.
2.
Ensure that the correct device is connected.
•
The
Matched Thru
verification standard is the device with one scribe line.
•
The
Mismatched Thru
verification standard is the device with two scribe lines.
3.
Enter the serial number of the Matched Thru.
4.
When ready to proceed, click
OK
. The
Connect Mismatched Thru Male Connector
dialog box appears.
5.
Connect the Matched Thru male connector to VNA Port 1 with the label facing up.
Note
The verification devices must be connected to the
0.8 mm Connector Coupler Ports
and not to the
VNA Ports
. Use the torque wrench supplied with the verification kit when tightening the device
connections.
Figure 2-39.
Enter Serial Number Thru Dialog
Note
The verification standards must be connected to the
0.8 mm Coupler Ports
and not to the
VNA
ports
.
Figure 2-40.
Connect the Mismatched Thru Dialog