4-7 Low-Band Calibration
Using the 2300-580-R Software with VectorStar ME7838G
4-18
PN: 10410-00327 Rev. D
3659 0.8 mm Cal/Ver Kit & 2300-580-R PVS UG
4-7
Low-Band Calibration
During this procedure, the low-band portion of the VNA is calibrated with a full 12-term SOLT
(Short-Open-Load-Thru) calibration. In
Section 4-8 “High-Band Calibration” on page 4-22
, the high-band
portion of the VNA is calibrated with a full 12-term SSST (Short-Short-Short-Thru) calibration.
Procedure
1.
The
0.8 mm Verification Install Adapter
dialog box appears.
a.
Install the 33.8G50 adapters on MMW-Modules Port 1 and Port 2.
Figure 4-23.
Anritsu 0.8 mm Verification Dialog