4-6 Application Interface – Setup Menu Tab
Using the PVS Software with the ME7828A
4-10
PN: 10410-00784 Rev. A
3656C W1 Cal/Ver Kit & 2300-616-R PVS UG
4-6
Application Interface – Setup Menu Tab
The
PVS Application Interface
dialog box displays three tabs with configuration, control, and serial number
information. Each tabbed dialog box contains buttons to control program operations as outlined in the
following sections:
•
The
Setup Menu
tab is described in this section.
•
The
Main Menu
tab display is described in
Section 4-9 “Application Interface – Main Menu Tab”
•
The
Serial Number
tab is described in
Section “Serial Number Tab Functions” on page 4-26
.
The following procedure continues the program setup from the previous section:
1.
In the
Calibration Kit Type
area, select
W Coax
.
2.
In the
Serial Number (5 Digits)
area, enter the serial number for the 3656C Calibration/Verification Kit. At
least five (5) digits are required.
3.
In the
Load Cal-Kit Coefficients to VNA
area, select
Yes
.
4.
The
Install Calibration-Kit Coefficients to the VNA
dialog box appears.
Figure 4-15.
ME7828A Application Interface – Setup Menu Tab
Cal-Kit
W1 Coax Model 3656C