Test Records
S312D MM
PN: 10580-00195 Revision B
A-3
Spectrum Analyzer Frequency Accuracy
Use the first row for the VNA Frequency Accuracy test. If the test fails, then use the second row for the
repeated test.
Spectrum Analyzer Phase Noise Verification
S312D
Firmware Revision:
________________
Operator:
________________
Date:
______________
Serial Number:
________________
Options:
___________________________________________________
Table A-3.
Spectrum Analyzer Frequency Accuracy
Measured Value
Specifications
Hz
1000 MHz ± 2 kHz
Hz
1000 MHz ± 2 kHz
Table A-4.
Spectrum Analyzer Phase Noise Verification
Reading
Delta 2
Delta 3
Specification
1
dBc
dBc
—
2
dBc
dBc
—
3
dBc
dBc
—
4
dBc
dBc
—
5
dBc
dBc
—
Average
dBc
dBc
—
Average – 30 dB
dBc/Hz
dBc/Hz
≤
–75 dBc/Hz