version 2.0 rev 18 Feb 2021
39
2.6.2.9 Rolling Shutter Global Clear Internal (Non-Overlap Mode)- Zyla 4.2 only
Rolling Shutter Global Clear
is an extended Rolling Shutter Mode. If enabled, the sensor rows are held in a
global clear state to ensure there is no charge build-up on the sensor. When the camera is taken out of the
Global Clear state, the exposure period begins for all rows in the sensor. When the required exposure period for
Row 1 has completed, a signal frame rolling shutter readout phase begins. This means all rows have the same
exposure start time but different exposure end times.
When the frame has been read out completely, the camera is returned to the Global Clear state until the user-
defined cycle time is achieved or external trigger is detected (if in an external triggering mode).
Rolling Shutter Global Clear mode is best used in conjunction with illumination controlled by the Fire ALL signal.
Figure 23: Rolling Shutter Global Clear- Internal (Non-Overlap Mode) - Zyla 4.2
Table 18:
Rolling Shutter Global Clear Internal (Non Overlap Mode) Timing Parameters - Zyla 4.2
Parameter
Minimum
Maximum
Exposure*
1 Frame + 1 Row
30s
Cycle Time (1/Frame Rate)
Ex 1 Frame + 5 Rows
20,000s
Acquisition Start Delay
-
2 Rows
*The exposure time reported is the exposure time for ROW 1 and the period for which FIRE ALL is HIGH.
Содержание Zyla sCMOS 4.2 PLUS
Страница 76: ...version 2 0 rev 18 Feb 2021 76...