Andon KD-927 Скачать руководство пользователя страница 1

DN:KD-927-SMSY01 V2.0                                                                    Operation Guide 

 

History Revision 

NO 

Revision 

date 

Revision Description 

Version

Drafter  Approver 

01  2013/09/09      Originated 

V1.0 

王成正

 

 

 

 

02  2014/9/5 

Update the address 

of EUROPEAN 

REPRESENTATION

V2.0  

费凡

 

   

张飞

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Содержание KD-927

Страница 1: ...927 SMSY01 V2 0 Operation Guide History Revision NO Revision date Revision Description Version Drafter Approver 01 2013 09 09 Originated V1 0 02 2014 9 5 Update the address of EUROPEAN REPRESENTATION...

Страница 2: ...DN KD 927 SMSY01 V2 0 Operation Guide Operation Guide V2 0 Product project name Blood Pressure Monitor Model name KD 927 Project number 2012 066 RJ17 Drafted by Date Reviewed by Date Approved by Date...

Страница 3: ...ITHIUM BATTERY CHARGING 6 2 CLOCK AND DATE ADJUSTMENT AND BLUETOOTH SET 6 3 CONNECTING THE CUFF TO THE MONITOR 8 4 APPLYING THE CUFF 9 5 BODY POSTURE DURING MEASUREMENT 9 6 TAKING YOUR BLOOD PRESSURE...

Страница 4: ...est value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most people are aw...

Страница 5: ...show blood pressure and pulse rate The most recent 2 60 measurements can be stored in the memory with date and time stamp The Electronic Sphygmomanometers corresponds to the below standards IEC 60601...

Страница 6: ...RH 14 Environmental temperature for storage and transport 20 55 4 131 15 Environmental humidity for storage and transport 90 RH 16 Environmental pressure 80kPa 105kPa 17 Battery life Approx 50 times...

Страница 7: ...be accurate it s suggested that you consult with your physician for accurate assessment There are 2 conditions under which the signal of IHB will be displayed 1 The coefficient of variation CV of pul...

Страница 8: ...changes or modification not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment SETUP AND OPERATING PROCEDURES 1 LITHIUM BATTERY CHARGIN...

Страница 9: ...ressing the button M2 the number will increase fast press the button M1 to decrease the number Keep on pressing the button M1 the number will decrease fast Picture 2 2 Picture 2 3 Picture 2 4 Picture...

Страница 10: ...our format and 12 hour format Table 1 24 hour format 12 hour format 24 hour format 12 hour format 0 00 12 00 AM 12 00 12 00 PM 1 00 1 00 AM 13 00 1 00 PM 2 00 2 00 AM 14 00 2 00 PM 3 00 3 00 AM 15 00...

Страница 11: ...er to the cuff circumference range in SPECIFICATIONS to make sure that the appropriate cuff is used 2 Measuring on same arm each time 3 Do not move your arm body or the monitor and do not move the rub...

Страница 12: ...can press the button M1 or M2 to go to the memory mode If you press M1 button you will scan user 1 memory If you press M2 button you will scan user 2 memory c If you no press any button the device wil...

Страница 13: ...e most frequent error made when measuring blood pressure is taking the measurement when not at rest which means that both the systolic and the diastolic blood pressures are incorrect in this case Duri...

Страница 14: ...current user See picture 7 6 If no result stored from 18 o clock to 20 o clock in last 7 days LCD will display 0 for blood pressure and pulse rate See picture 7 7 At this time you can go forward by p...

Страница 15: ...you press the button M1 you can press the button START STOP to turn off the monitor manually Picture 7 11 Note Above assume scan user 1 memory if you want to scan user 2 memory you need to press the...

Страница 16: ...t Mode and bluetooth Mode connect the monitor to your computer by USB cable the device display PC and data and time See picture 9 1 Picture 9 1 c Press the M1 button or M2 button to change user Press...

Страница 17: ...he monitor h After transmission the data will not be deleted from the device i Please turn off the monitor and disconnect it with your computer after downloading Note 1 Don t turn off the monitor whil...

Страница 18: ...ns The technical alarm condition outside the rated range is preset in the factory and cannot be adjusted or inactivated This alarm condition is assigned as low priority according to IEC 60601 1 8 The...

Страница 19: ...in LCD shows Er 4 Pneumatic system leakage or cuff is too loose during inflation LCD shows Er 5 Cuff pressure above 300mmHg Measure again after five minutes If the monitor is still abnormal please con...

Страница 20: ...cs for a minimum of 10 000 measurements or three years and the cuff integrity is maintained after 1 000 open close cycles of the closure 9 It is recommended the cuff should be disinfected 2 times ever...

Страница 21: ...se in the electromagnetic environment specified below The customer or the user of the KD 927 should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment...

Страница 22: ...t Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage d...

Страница 23: ...eparation distance 80 MHz to 800 MHz 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended sepa...

Страница 24: ...etic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the KD 927 as recommended below according to the maximum output power of th...