Thank you for purchasing the Ambulatory Blood Pressure Monitor TM2441-EX. The Instruction Manual that
accompanies the product has been corrected and had content added. Please read it in conjunction with the
details shown below.
The following additions have been made.
Addition
When the AUTO switch is
"ON"
, environmental data are only saved during automatic
measurement operations.
Caution
Environmental data are saved for up to 12 days, but in this case, it will take about
50 minutes to import the data into the ABPM Data Manager.
The following changes have been made.
Wrong
Correct
An initial pressurization value can be set beforehand.
160, 180, 210, 240, 270, AUTO [mmHg]
If AUTO is specified to the initial pressurization value, the
pressurization value is selected automatically. Factory
setting is 180 mmHg. Refer to
"8.2.3. Initial
Pressurization Value"
for information on how to select an
initial pressurization value.
The initial pressurization value for automatic blood pressure
measurement is AUTO.
In the case of AUTO, the appropriate pressurization value
is set automatically.
Ambulatory Blood Pressure Monitor TM2441-EX,
Instruction Manual, Corrections and Additions
Instruction manual Page
28 6.1.
Automatic Blood Pressure Measurement (A-BPM)
6.1.4
.
Saving Environmental Data
Instruction Manual Page
27 6.1
. Automatic Blood Pressure Measurement (A-BPM)