Thank you for purchasing the Ambulatory Blood Pressure Monitor TM2440-EX. The Instruction Manual that 

accompanies the product has been corrected and had content added. Please read it in conjunction with the 

details shown below. 

 

 

 

The following changes have been made. 

Wrong 

Correct 

The initial pressurization value is set to 180 mmHg at the 
factory. 

The initial pressurization value for automatic blood pressure 
measurement is AUTO. 
In the case of AUTO, the appropriate pressurization value 
is set automatically. 

 

 

 

 

The following changes have been made. 

Wrong 

Correct 

The initial pressurization value is set to 180 mmHg at the 
factory. 

The initial pressurization value setting is AUTO 
(appropriate pressurization value set automatically) when 
the

     

  mark is displayed and 180 mmHg when the

 

 

mark is hidden. 

 

 

 

 

The following is added after 

"

A-BPM can use only while the automated measurement can be performed. 

 

Addition 

 

Note

 

The 

Start time

 and 

Operation time

 are initially set calculating from the 

time  at  which  the 

mark  was  displayed.  When  the  next 

automatic measurement is made, this needs to be set again. 

 

 

Ambulatory Blood Pressure Monitor TM2440-EX,   

Instruction Manual, Corrections and Additions

 

Instruction manual Page 

22

 

6.1

. Automatic Blood Pressure Measurement (A-BPM) 

Instruction manual Page 

37

 

8.2.3. 

Initial Pressurization Value 

Instruction Manual Page 

37

 

8.3

. A-BPM Preset Programs 