1
1WMPD4004497
Thank you for purchasing the Ambulatory Blood Pressure Monitor TM2440-EX. The Instruction Manual that
accompanies the product has been corrected and had content added. Please read it in conjunction with the
details shown below.
The following changes have been made.
Wrong
Correct
The initial pressurization value is set to 180 mmHg at the
factory.
The initial pressurization value for automatic blood pressure
measurement is AUTO.
In the case of AUTO, the appropriate pressurization value
is set automatically.
The following changes have been made.
Wrong
Correct
The initial pressurization value is set to 180 mmHg at the
factory.
The initial pressurization value setting is AUTO
(appropriate pressurization value set automatically) when
the
mark is displayed and 180 mmHg when the
mark is hidden.
The following is added after
"
A-BPM can use only while the automated measurement can be performed.
"
Addition
Note
The
Start time
and
Operation time
are initially set calculating from the
time at which the
mark was displayed. When the next
automatic measurement is made, this needs to be set again.
Ambulatory Blood Pressure Monitor TM2440-EX,
Instruction Manual, Corrections and Additions
Instruction manual Page
22
6.1
. Automatic Blood Pressure Measurement (A-BPM)
Instruction manual Page
37
8.2.3.
Initial Pressurization Value
Instruction Manual Page
37
8.3
. A-BPM Preset Programs