background image

 

9

5-6. Weighing range for the EW-

i

 series 

 

‰

 The EW-

i

 series have three weighing ranges, and the display shows which range 

the weight value belongs to with the mark R1, R2 or R3. 

‰

  There is a function setting to select how the weighing range changes. 

‰

 Select from automatic range (

rng

 

1

), manual range (

rng

 

0

) or fixed range (

rng

 

2

 

to 

4

). 

 

Operation 

 

Function setting

Operation 

rng 1

 

Automatic range 

‰

  When the weight value exceeds the maximum value of 

the range, the weighing range changes automatically 
from a lower to a higher weighing range. 

‰

  When there is nothing on the weighing pan and the 

display shows zero with the ZERO mark, the weighing 
range changes from a higher to the lowest range. 

‰

  When the    RE-ZERO    key is pressed in a higher range, 

the balance will be tared to zero and the weighing range 
changes to the lowest range. 

 

rng 0

 

Manual range 

‰

  Press the  SAMPLE  key when the display shows a 

weight value (neither counting nor % display). The 
weighing range changes to a higher range at any load.   

‰

  Press the    SAMPLE    key to change from a higher to the 

lowest range, when there is nothing on the weighing pan 
and the display shows zero with the ZERO mark.   

‰

  When the    RE-ZERO    key is pressed in a higher range, 

the balance will be tared to zero and the weighing 
range changes to the lowest range. If the weight of the 
object is not more than 2% of the weighing capacity, the 

 RE-ZERO  key doesn’t tare, but zeroes the balance 

and the weighing range doesn’t change.   

 

Press the    SAMPLE    key to change the weighing range 

at zero display. 

 

rng 2 

to

 4

 

Fixed range 

(

*

)

 

‰

  The weighing range is fixed. Set the function to the 

weighing range according to the purpose. 

 

 

(

*

)

 

This function will not be available for some of legally certified models. 

 
 
 
 
 
 

Содержание EK-1200i

Страница 1: ...x 1WMPD4000465B ...

Страница 2: ...ent has been tested and has been found to comply with the limits of Class A computing device pursuant to Subpart J of Part 15 of FCC rules These rules are designed to provide reasonable protection against interference when the equipment is operated in a commercial environment If this unit is operated in a residential area it may cause some interference and under these circumstances the user would ...

Страница 3: ...ounting mode pcs 10 5 8 Percent mode 11 6 COMPARATOR 12 6 1 Setting example 12 7 CALIBRATION 14 7 1 Calibration using a weight 14 7 2 Gravity acceleration correction 16 8 FUNCTIONS 17 8 1 Key operation 17 8 2 Entering the function setting mode 17 8 3 Setting example 18 8 4 Storing weighing units 19 8 5 Function list 20 9 RS 232C SERIAL INTERFACE 22 9 1 Interface specifications 22 9 2 Data format 2...

Страница 4: ...ssembly 31 11 3 OP 09 Rechargeable battery pack Ni MH 32 11 4 OP 12 Carrying case 32 12 MAINTENANCE 33 12 1 Notes on maintenance 33 12 2 Error codes 33 13 SPECIFICATIONS 35 13 1 EK i series 35 13 2 EW i series 36 13 3 Other weighing units 37 13 4 Dimensions 38 GRAVITY ACCELERATION MAP 39 ...

Страница 5: ... use in a dimly lighted place The standard RS 232C serial interface can be connected to a printer or personal computer Using the serial interface Good Laboratory Practice GLP data can be output The optional rechargeable battery pack OP 09 is easy to install for cordless operation 2 UNPACKING When unpacking check whether all of the following items are included Main unit EW 1500i EW 12Ki EK 600i EK ...

Страница 6: ...es the weighing range when rng 0 is selected pcs mode Enters the sample unit weight storing mode mode Enters the 100 weight storing mode Outputs the weight data to a printer or computer or enters a unit weight 100 weight or other setting values to the balance Switches the weighing unit the weighing mode Clears the display to zero ...

Страница 7: ...ce where combustible or corrosive gases may exist Allow the balance to reach equilibrium with the ambient temperature before use Switch the power ON at least half an hour before use so that the balance can warm up When the balance is installed for the first time or the balance has been moved carry out calibration as described in 7 CALIBRATION 4 2 Power source For the power source the AC adapter or...

Страница 8: ...ere is a load beyond this range the balance will be tared to zero and the NET mark and the ZERO mark turn on 2 Pressing the ON OFF key again and the power will be switched OFF Auto power off function It is possible to have the power automatically switched OFF if zero is displayed for approximately 5 minutes See 8 5 Function list and set the function poff 5 2 LCD backlight The LCD backlight will tu...

Страница 9: ...s are not available for higher or lower capacity models For details see 13 SPECIFICATIONS Note It is possible to store only the units that will be actually used from the units available It is also possible to specify the display unit that will be shown first when the power is switched ON For details see 8 4 Storing weighing units Conversion table Units Name Conversion to gram oz Ounce avoir 28 349...

Страница 10: ...alance if the weight is within 2 of the weighing capacity around the power on zero point The ZERO mark turns on When the weight exceeds 2 of the weighing capacity it will be subtracted to zero as a tare weight In this case the ZERO and NET marks turn on Precautions during operation Make sure that the stability mark is on whenever reading or storing a value Do not press the keys with a sharp object...

Страница 11: ...l be tared to zero and the weighing range changes to the lowest range rng 0 Manual range Press the SAMPLE key when the display shows a weight value neither counting nor display The weighing range changes to a higher range at any load Press the SAMPLE key to change from a higher to the lowest range when there is nothing on the weighing pan and the display shows zero with the ZERO mark When the RE Z...

Страница 12: ...mple The balance is set to count objects with this unit weight When a unit weight is too light to store the display shows lo for a moment and returns to the former display When the sample weight is light and the counting error could become large the balance will prompt you to use a larger number of samples Place the displayed number of samples on the pan and press the PRINT key to calculate and st...

Страница 13: ...samples as displayed 5 8 Percent mode Displays the weighing value in percentage compared to the reference 100 weight Selecting the percent mode 1 Press the MODE key to select percent Storing the reference 100 weight 2 Press the SAMPLE key to enter the reference weight storing mode 3 Press the RE ZERO key to display 100 0 4 Place the sample to be set as the reference weight on the pan 5 Press the P...

Страница 14: ...four display divisions The upper limit and lower limit numerical values are common to each of the weighing counting and percent mode An example for the EK 1200i 2000i 3000i is as follows Upper limit value 001010 101 0g 1010pcs 101 0 Lower limit value 000990 99 0g 990pcs 99 0 6 1 Setting example This example will be Compares plus data except those near zero Selecting a comparison mode 1 Press and h...

Страница 15: ...ep 7 With Cp lo displayed press the PRINT key Enter the lower limit value using the following keys SAMPLE key To select the digit blinking to change RE ZERO key To set the value of the digit selected Hold down the key to switch the sign and see step 6 PRINT key To store the value and proceed to the next step MODE key To cancel the value and proceed to the next step 8 Press the PRINT key Unit appea...

Страница 16: ...ys Cal 0 To change the calibration weight value proceed to step 4 To use the calibration weight value in the balance memory proceed to step 5 4 Press the SAMPLE key The display shows the calibration weight value in gram that is stored in the balance Use the following keys to change the value SAMPLE key To select the digit blinking to change RE ZERO key To set the value of the digit selected PRINT ...

Страница 17: ...t Do not touch the pan during weighing 7 end appears Remove the weight from the pan and press the CAL switch or MODE key to return to the weighing mode Note The value set in step 4 is stored in memory even after the power is switched off If the balance is to be moved to other places set the gravity acceleration value of the area where the calibration using a weight is to be done and calibrate the ...

Страница 18: ...tion CAL switch until Cal appears and release the switch 2 The balance displays Cal 0 3 Press the RE ZERO key The display shows the gravity acceleration value stored in the balance Use the following keys to change the value SAMPLE key To select the digit blinking to change RE ZERO key To set the value of the digit selected PRINT key To store the value and return to step 2 MODE key To cancel the va...

Страница 19: ...or selection See Function list Cancels the operation and turns off the balance Item key Selects a class and an item In the weighing mode press and hold the key to enter the function setting mode Enter key Proceeds to the selected class Stores the settings per class and goes to the next class Parameter key Selects a parameter Cancel key Cancels the operation and goes to the next class or returns to...

Страница 20: ...ey to display poff 1 4 Press the SAMPLE key several times to display aCai 1 5 Press the RE ZERO key to select aCai 0 6 Press the PRINT key to store the parameters Cp Hi appears after end 7 Press the MODE key to return to the weighing mode func poff 0 poff 1 aCai 1 Press and hold To confirm To store aCai 0 end Cp Hi Returns to the weighing mode Each pressing switches the parameter Each pressing swi...

Страница 21: ... key 4 Press the SAMPLE key to select a weighing unit 5 Press the RE ZERO key to store the weighing unit 6 Repeat steps 4 and 5 to store all weighing units to be used 7 Press the PRINT key id appears after end 8 Press the MODE key to return to the weighing mode Note When the balance is switched on it starts with the unit that was stored first at step 5 func Unit Unit Press and hold Unit end id Ret...

Страница 22: ...e data 3 Compares plus data 4d 4 Compare stable plus data 4d 5 Compares data 4d or 4d Cp Comparator mode 6 Compares stable data 4d or 4d Conditions to compare d the minimum display division 0 Buzzer does not sound 1 Buzzer sounds at LO 2 Buzzer sounds at OK 3 Buzzer sounds at OK and LO 4 Buzzer sounds at HI 5 Buzzer sounds at HI and LO 6 Buzzer sounds at HI and OK bep Buzzer output 7 Buzzer sounds...

Страница 23: ...onds 2 Turns off after 10 seconds 3 Turns off after 30 seconds 4 Turns off after 60 seconds func ltUp LCD Backlight control 5 Always on To control how the LCD backlight turns off Weight change or key operation will turn the backlight on CpHi Comparator upper limit Setting the upper limit value Cplo Comparator lower limit Setting the lower limit value See 6 COMPARATOR Unit Weighing units to be disp...

Страница 24: ...aight type to connect with a computer Windows Communication Tools Software WinCT to communicate with a computer is provided as freeware Visit the A D website to download WinCT 9 1 Interface specifications Transmission system EIA RS 232C Transmission form Asynchronous bi directional half duplex Data format Baud rate 1200 2400 4800 9600 bps Data 7 bits parity 1 bit even or odd or 8 bits non parity S...

Страница 25: ...e c t Weight data carat mom Weight data momme d w t Weight data penny weight GN Weight data grain t l Weight data tael The terminator is always CRLF Example of output data Weight data gram S T 0 0 1 2 3 4 5 g CR LF Counting data Q T 0 0 0 1 2 3 4 5 P C CR LF Percentage data S T 0 0 0 1 2 3 4 CR LF Out of range gram O L 9 9 9 9 9 9 9 g CR LF Out of range pcs O L 9 9 9 9 9 9 9 9 P C CR LF 9 3 Data o...

Страница 26: ...o print mode B Set the function prt 3 The balance transmits the weight data when the display is stable the stability mark is on and the data is greater than 4d of the lowest range for EW i series or less than 4d The next output can be obtained after the display returns between 4d and 4d 9 4 Command mode In the command mode the balance is controlled by commands that come from the personal computer ...

Страница 27: ...ber using the following keys SAMPLE key To select the digit blinking to change RE ZERO key To set the character of the digit selected See the table below for the display character set PRINT key To store the value and proceed to the next step MODE key To cancel the value and proceed to the next step 4 When the above operation has completed func appears after end 5 Press the MODE key to return to th...

Страница 28: ...the calibration has been completed 3 glp is displayed and calibration report is output 4 end appears again Remove the weight and press the MODE key to return to the weighing mode AD 8121 format info 1 General format info 2 Space ASCII 20h CR Carriage return ASCII 0Dh LF Line feed ASCII 0Dh end glp To output calibration report Operation of Calibration See 7 CALIBRATION end Returns to the weighing m...

Страница 29: ...ation test weight according to the procedure step 4 of 7 1 Calibration using a weight 4 With nothing on the pan press the PRINT key The zero point is measured and the weighed value is displayed for a few seconds Then the display shows the value of calibration test weight 5 Place a weight of the same value as displayed on the pan and press the PRINT key to weigh it The weighed value is displayed fo...

Страница 30: ... the weight data by pressing the PRINT key or selecting the auto print mode A D MODEL EK 1200i S N Q1234567 ID ABCDEF DATE 2011 11 14 TIME 03 15 40 CAL TEST EXT ACTUAL 0 0 g 1000 1 g TARGET 1000 0 g SIGNATURE A D CRLF MODEL EK 1200i CRLF S N Q1234567 CRLF ID ABCDEF CRLF DATE CRLF CRLF TIME CRLF CRLF CAL TEST EXT CRLF ACTUAL CRLF 0 0 g CRLF 1000 1 g CRLF TARGET CRLF 1000 0 g CRLF SIGNATURE CRLF CRL...

Страница 31: ...2011 11 17 START TIME 01 23 45 ST 000123 4 g ST 000234 5 g ST 000345 6 g ST 000456 7 g END TIME 01 25 43 SIGNATURE A D CRLF MODEL EK 1200i CRLF S N Q1234567 CRLF ID ABCDEF CRLF DATE CRLF CRLF START CRLF TIME CRLF CRLF CRLF ST 000123 4 g CRLF ST 000234 5 g CRLF ST 000345 6 g CRLF ST 000456 7 g CRLF END CRLF TIME CRLF CRLF SIGNATURE CRLF CRLF CRLF CRLF CRLF CRLF recend Returns to the weighing mode e...

Страница 32: ...e to sound a buzzer according to the comparison result See the function bep to set which result will make a buzzer sound The comparator function on off the comparison mode and comparator buzzer output can be selected using the function settings See the settings cp and bep OP 04 cannot be used together with OP 09 Output circuit Maximum rating The maximum rating of the relay output is as follows Max...

Страница 33: ... bottom of the balance large objects that are difficult to load on the weighing pan can be weighed in suspension and the specific gravity of objects may be measured OP 07 is available for use with the EK 6000i 12Ki 4100i 6100i and EW 12Ki balances only The calibration with a weight being hung on the hook is required for an accurate weighing OP 07 Installation Remove the cover on the bottom of the ...

Страница 34: ...ely stop using it and recharge the battery pack or use the AC adapter Charge the battery pack at a temperature between 0 C 32 F and 40 C 104 F Do not charge too long Overcharging will reduce the life of the batteries Be sure to charge the battery pack when using for the first time or if it has not been used for a long time more than one month Recharging two or three times may be needed to reach fu...

Страница 35: ...the weight sensor receives strong force upward Check if there is anything sandwiched around the weighing pan There is a possibility that the weight sensor itself may have a failure Unit 100 weight error The unit 100 weight is too light in the counting mode Sample quantity notice When sample weight is light and the counting error could become large the balance will prompt you to use a larger number...

Страница 36: ...arning to indicate that the output voltage of the AC adapter is too high Check if the AC adapter is correct Stability error Warning to indicate that the weight value is not stable and the balance cannot display it Prevent vibration and drafts Press the MODE key to return to the weighing mode Internal errors 2 3 4 or 6 The balance detects error state in the internal processing With nothing on the w...

Страница 37: ... 1 kg 1 1 kg 1 3 kg Calibration weight factory setting 120 g 200 g 300 g 600 g MODEL EK 1200i EK 2000i EK 3000i EK 6000i EK 12Ki Weight capacity 1200 g 2000 g 3000 g 6000 g 12 kg Min display d 0 1 g 0 1 g 0 1 g 1 g 1 g No of samples 5 10 25 50 or 100 pieces Max count 12 000 pcs 20 000 pcs 30 000 pcs 6000 pcs 12 000 pcs Min unit weight 0 1 g 0 1 g 0 1 g 1 g 1 g Min display 0 1 Min 100 weight 10 g 1...

Страница 38: ...ght factory setting 400 g 600 g 4000 g 6000 g In case of Umin 0 factory setting 13 2 EW i series MODEL EW 150i EW 1500i EW 12Ki Weight capacity 30 g 60 g 150 g 300 g 600 g 1500 g 3 kg 6 kg 12 kg Min display d 0 01g 0 02g 0 05g 0 1g 0 2g 0 5g 1g 2g 5g No of samples 5 10 25 50 or 100 pieces Max count 15 000 pcs 15 000 pcs 12 000 pcs Min unit weight 0 01 g 0 1 g 1 g Min display 0 1 Min 100 weight 1 g...

Страница 39: ... 228 lb Min display 0 0001 0 0001 0 001 0 001 Capacity 12 860 19 290 128 60 192 90 ozt Min display 0 001 0 001 0 01 0 01 Capacity 2000 0 3000 0 20000 30000 ct Min display 0 1 0 1 1 1 Capacity 106 665 160 000 1066 65 1600 00 mom Min display 0 005 0 005 0 05 0 05 Capacity 257 21 385 81 2572 1 3858 1 dwt Min display 0 01 0 01 0 1 0 1 Capacity 6173 0 9259 4 61730 92594 GN Min display 0 2 0 2 2 2 Capac...

Страница 40: ...38 13 4 Dimensions EK 120i EK 200i EK 300i EK 600i EK 1200i EK 2000i EK 3000i EK 6000i EK 12Ki EK 410i EK 610i EK 4100i EK 6100i EW 150i EW 1500i EW 12Ki ...

Страница 41: ...m s2 Ottawa Paris 9 807 m s2 9 810 m s2 Copenhagen 9 816 m s2 Rio de Janeiro 9 788 m s2 Cyprus 9 797 m s2 Rome 9 803 m s2 Frankfurt 9 811 m s2 San Francisco 9 800 m s2 Glasgow Havana 9 816 m s2 9 788 m s2 Singapore Stockholm 9 780 m s2 9 819 m s2 Helsinki 9 819 m s2 Sydney 9 796 m s2 Jakarta 9 781 m s2 Taichung 9 789 m s2 Kolkata Calcutta 9 788 m s2 Taipei 9 790 m s2 Kuwait Lisbon London Greenwich...

Страница 42: ...40 World map ...

Страница 43: ...nsburg Deutschland Telefon 49 0 4102 459230 Telefax 49 0 4102 459231 A D Australasia Pty Ltd 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D KOREA Limited 한국에이 엔 디 주 대한민국 서울시 영등포구 여의도동 36 2 맨하탄 빌딩 8층 8th Floor Manhattan Bldg 36 2 Yoido dong Youngdeungpo ku Seoul KOREA 전화 82 2 780 4101 팩스 82 2 782 4280 A D RUS CO LTD Компания ЭЙ энд ДИ РУС 1213...

Отзывы: