AD-1683 

DC Static Eliminator 

          Instruction Manual 

 
 

© 2004 A&D Company Ltd. All rights reserved. 

 

No part of this publication may be reproduced, 
transmitted, transcribed, or translated into any language 
in any form by any means without the written permission 
of A&D Company Ltd. 
 
The contents of this manual and the specifications of the 
instrument covered by this manual are subject to change 
for improvement without notice. 

 

 

 

 
 
 
 

To Use the AD-1683 Safely

 

Thank you for your purchasing the AD-1683 DC static eliminator. 

Please read this manual completely before using the AD-1683. 

 

Warning Definition 

The warning described in this manual has the following meaning: 

 

        Caution 

Important information to alert you to a situation 

that may cause injury and/or damage to your 

property if instructions are not followed. 

When operating this instrument, be sure to observe the 

following safety precautions 

Caution 

 

Do not open the case to repair the AD-1683. Only qualified 

personnel can do that. Attempting repairs yourself may cause 

damage to the AD-1683. Damage caused by attempting to do 

the repair yourself will void the warranty.

 

 Do not use the AD-1683 in a closed environment. Ozone 

that is harmful to the human body may stray into the air. 

Ventilate well. 

 The AD-1683 is not explosion-protected. Do not use the 

instrument where flammable solvents or dust exits. 

 Do not touch the discharge electrode pins. It may cause an 

electrical shock. 

 Use the specified AC adapter only. Please confirm that the 

AC adapter type is correct for your local voltage and 

receptacle type. If an AC adapter other than specified is 

used, static electricity may not be eliminated or it may 

cause an electrical shock. (*1) 

 Be sure to insert the AC adapter into a grounding-type 

electrical outlet and earth ground the grounding cable. 

Otherwise, static electricity may not be eliminated or it may 

cause an electrical shock. 

*1: Use the AC adapter identification label to identify the AC adapter. 

Caution 

  Do not keep the AC adapter connected to the electrical 

outlet for a long period of time. Dust that has collected on 

the outlet may cause a fire. 

 When relocating the AD-1683, remove the AC to avoid a 

possible electrical shock. 

 

Precautions on Usage

 

Caution 

 

If the AD-1683 malfunctions, stop using it immediately. Place 

an “Out of Order” label on it or move the instrument to a 

place where it can not be used by mistake. It is very 

dangerous to continue using the instrument under such 

conditions. Ask the nearest dealer for services. 

 Do not face the front of the AD-1683 directly toward an 

electronic balance. Ionized air from the AD-1683 may affect 

the measurement and cause a measurement error. 

 When used with a balance of other manufacturers, static 

elimination can not be guaranteed.   

 Do not place the sample closer to the AD-1683 than specified. 

The sample may become charged. 

 Do not place any obstacle between the AD-1683 and the 

sample.  

 The stand can be attached to the workbench. The stand has 

three holes with a diameter of 6 mm in 42.5 mm interval for 

that purpose.   

(To secure the stand: Bend the side plates outward to remove 

the AD-1683, using much care not to separate the upper and 

lower cases. Attach the stand to the workbench using the 

screws. Replace the AD-1683 on the stand.) 

 

 

 

 

 

1. Introduction 

This manual describes how the AD-1683 works and how to get the 

most out of it in terms of performance. Please read this manual 

completely before using the AD-1683. 

 

2. Features 

The AD-1683 generates bipolar ions continuously by DC corona 

discharge separately from positive and negative discharge 

electrodes and directs the ionized air onto the charged body to 

eliminate static electricity. The generated ions are well-balanced 

in polarity and can eliminate static electricity regardless of the 

polarity of the charged body. (See Fig.3.) 

 

Static electricity 

Generally, insulators such as plastics are prone to be charged 

with static electricity when the ambient relative humidity is 

below 45%RH. This may affect weighing and cause a 

measurement error of several milligrams.   

The AD-1683 can eliminate static electricity very effectively.   

 

3. Part Names 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 

Note: Please confirm that the AC adapter type is correct for your 

local voltage and receptacle type. 

 

To prevent electrical shock, insert the POWER cable into a 

grounding-type outlet and earth ground the grounding cable. 

 

4. How to Use the AD-1683 

4-1. Installing the AD-1683 

Choose the installation site so 

that the AD-1683 can be placed 

beside the electronic balance 

and a space wide enough to 

place a sample (10 cm to 30 cm 

from the front of the AD-1683) 

can be secured. Place the 

sample in front of the AD-1683 

inside the specified area and 

perform static elimination. Then, 

weigh the sample using the 

electronic balance. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

4-2. Turning the power on 

Insert the AC adapter plug to the AC adapter jack located on the 

rear of the AD-1683. Insert the AD adapter into an electrical 

outlet. The red LED at the center of the AD-1683 illuminates to 

indicate that static elimination is available. 

Note: Please confirm that the AC adapter type is correct for 

your local voltage and receptacle type.. 

 

4-3. Finishing the use 

Remove the AD adapter from the electrical outlet. 

 

5. Maintenance 

5-1.Cleaning 

Caution 

Remove the AC adapter from the AD-1683 before cleaning. 

 

When used for a long period of 

time, dust may collect around the 

discharge electrode pins and 

static elimination performance will 

be lowered. To maintain the 

elimination performance, clean the 

AD-1683 periodically (once a 

week). Use a cotton swab to clean 

the electrode pins. 

 
 
 
 
 
 
 

5-2. Replacing the discharge electrode pin 

When the elimination performance 

remains low even after cleaning 

the discharge electrode pins, 

replace the discharge electrode 

pins with new ones. 

Remove the AC adapter from the 

receptacle and the AD-1683. 

Using a pair of pliers, pull the 

discharge electrode pin out.   

Insert a new pin, using much care 

not to bend the pin tip. Be sure to 

insert the pin till stopped. 

 
Replacement discharge electrode pins (Sold separately):   
 AX-054016580-S 
 

6. Specifications 

Static elimination method 

DC corona discharge (Bipolar) 

Elimination range 

(See Fig. 2) 

 

Distance: Approx. 10 cm to 30 cm from the electrode 

 

Area : 30x30 cm from the electrode 

Elimination performance 

When charged 5 kV 

Distance 10cm

20cm 30cm

50cm

Elimination time 

1 sec

4 sec 15 sec 70 sec

Ambient temperature and humidity  0°C to 40°C, 80%RH or less 
Ozone concentration 

Positive discharge pin: 0.07PPM 

 

Negative discharge pin: 0.25PPM 

 

(Measured at a distance of 1 cm from each discharge pin.) 

Power supply 

AC adapter. Please confirm that the AC adapter type 
is correct for your local voltage and receptacle type. 

Power consumption 

Approx. 1.5 VA 

Discharge electrode pin 

Tungsten (

φ

0.9x17 mm), Life Approx. 10000 hours 

Mass 

Approx. 300 g (including the stand) 

 
 
 
 
 
 
 
 
 

External Dimensions 

 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 

 
 

 