Specification
39
10
Specification
10.1
Technical data
10.1.1
qTOWER
3
auto
Designation/type
qTOWER
3
auto
qTOWER
3
84 auto
Dimensions
(width x height x depth)
310 x 479 x 365 mm (drawer moved in)
310 x 479 x
477
mm (drawer moved out)
Traveling distance of the
drawer
112 mm
Weight
approx. 27 kg
Position tolerance of the
sample carrier when the
drawer is moved out
± 0.15 mm
Sample carrier
Verified for automatable processes, full-skirted
Sample block
Silver
(gold-plated)
Aluminum
(special alloy)
Sample capacity
96
384
Sample volume
10 –
100 μl
2 –
30 μl
(5 –
20 μl recommended)
Heating
max. 8 °C/s
max. 4 °C/s
Cooling
max. 5.5 °C/s
max. 2 °C/s
Temperature setting range
4 °C – 99 °C
Temperature uniformity
55 °C ± 0.15 °C after 15 s
Temperature control accu-
racy
± 0.1 °C
Heating rate adjustment
min. 0.1 °C/s
Temperature uniformity
(after 15 seconds)
± 0.15 °C at 55 °C
± 0.25 °C at 72 °C
± 0.50 °C at 95 °C
Temperature increments
min. 0.1 °C/cycle
Time increments
min. 1 s/cycle
Lid temperature range
30 °C – 110 °C
Contact force
approx. 300 N, automated
General
data
Thermal block
Technical characteristics
Heated lid
Technical characteristics