Palm CO

Palm CO

Palm CO

 

 

 

Carbon 

Carbon 

Carbon 

 

 

 

Monoxide

Monoxide

Monoxide

 

 

 

Analyzer

Analyzer

Analyzer

 

 

 

Assembled under a certified 

Quality Management System  



MDD/93/42 EEC,  Annex II  



ISO 13485:2003 



ISO 9001:2008

 

Instructions 

for Use 

Содержание Palm Co

Страница 1: ...b www aii1 com This manual may not be reproduced in whole or in part with out the prior written consent of Analytical Industries Inc Palm CO Palm CO Palm CO Carbon Carbon Carbon Monoxide Monoxide Mono...

Страница 2: ...y jeopardize the well being of the user Users must read the following statements as they are essential to re ducing the risk of use error due to ergonomic features of the device or the environment in...

Страница 3: ...nstable or any malfunction is suspected DO NOT use while ERR or bA messages are displayed on the LCD DO NOT use near equipment capable of emitting high levels of electro magnetic radiation EMI or radi...

Страница 4: ...arbon monoxide balance air from local sources The device should NEVER be bump tested with a gas containing more than 25 PPM carbon monoxide concentration 4 3 3 1 Bump Test Procedure 1 Set up sampling...

Страница 5: ...PM CO with the balance air throughout the entire span calibration procedure 2 ALLOW the reading on the analyzer s display to stabilize which can take up to two minutes 3 PRESS the CAL key for three 3...

Страница 6: ...rectly on the PCB assembly FIG 1 5 Operation 5 1 Principle of Operation The device utilizes an electrochemical carbon monoxide sensor intended to measure carbon monoxide levels in gas streams Carbon m...

Страница 7: ...re the wires are aligned with the registration slot FIG 6 and gen tly press down on the rear of the sensor until the sensor is seated 6 Connect the CO sensor reverse step 3 7 Reassemble the device rev...

Страница 8: ...reme tempera ture pressure or humidity 7 Spare Parts Accessories 13 FIG 8 FIG 7 8 Troubleshooting If the recommended corrective action does not resolve the problem return the device to the factory for...

Страница 9: ...ical Industries Inc shall not be liable for losses or damages of any kind loss of use of the analyzer incidental or consequential losses or damages damages resulting from alterations misuse abuse lack...

Страница 10: ...11 Specifications 16...

Страница 11: ...ibration in air 0 PPM 1 PPM 4 Reading before span calibration ___ PPM 5 PPM 5 Reading after span calibration ___ PPM 1 PPM 6 Reading in air after span calibration ___ PPM 1 PPM 7 Reading 24 hours afte...