Maintenance and care
52
Edition 06/2015
InnuPure C16
Only the use of ethanol or isopropanol containing disinfection agents are approved
for wiping disinfection. The use of disinfection agents with ingredients or concentra-
tions that differ from the recommended disinfection agents invalidates all warranty
claims regarding damage to the device or non-effectiveness.
If an InnuPure C16 has to be sent to Analytik Jena AG for service, a prior decon-
tamination must be carried out and documented (see the documents in the product
folder).
8.2
Changing fuses
CAUTION! Electric shock!
Disconnect the mains plug before changing fuses!
If the fuses in the InnuPure C16 are faulty you can replace them.
1.
Switch off the InnuPure C16 and disconnect the mains plug from the connection of the
InnuPure C16.
2.
Open the fuse holder by pulling its cover.
3.
Replace the faulty mains fuses. Only use 2 x T 4A; 250 V fuses.
4.
Close the fuse holder.
5.
Connect the InnuPure C16 to the mains.
6.
Switch on the InnuPure C16.
1
Mains switch
2
Fuses
3
Fuse holder
Figure 16 Open fuse holder