First commissioning
22
Edition 06/2015
InnuPure C16
5
First commissioning
5.1
Site requirements
5.1.1
Installation conditions
The following requirements are placed on the climatic conditions in the operating room of the
InnuPure C16:
Temperature range:
+
5°C
to +35 °C
Max. humidity:
80% at 30 °C
The following requirements are placed on the location of the InnuPure C16:
Do not locate the device directly near a door or window.
Place the device on a stable surface.
Do not locate the device near sources of electromagnetic interference.
Avoid direct sunlight and radiation from heaters on the device, if necessary provide
for air conditioning.
Never cover the front door or the ventilation slits of the device with other equipment
or installations!
Keep a safety distance of at least 5 cm from the rear of the device to other equip-
ment or walls!
5.1.2
Space requirement
The space requirement results from the device dimensions:
Closed device (W x D x H)
380 mm x 530 mm x 435 mm
Device with open front door (W x D x H)
380 mm x 920 mm x 435 mm
Sample tray (W x D x H)
200 mm x 250 mm x 83 mm
Priming station (W x D x H)
210 mm x 260 mm x 165 mm
The loaded sample tray is moved through the opening in the door into the device. The device
door must be opened for cleaning and maintenance purposes and must therefore be added
to the dimensions.
The ventilation slits on the rear of the device must remain unobstructed to ensure air circula-
tion.