12
Maintenance
It is essential to calibrate the gas sensors regularly to ensure their correct operation. It is
recommended that the system is calibrated prior to use.
Pressure, temperature and humidity sensors require less frequent calibration than the gas sensors.
Simple comparison checks with the Master Gauges during use will reveal faults with the pressure
sensors. In the event of a discrepancy, the sensors should be re-calibrated. It is expected that a
pressure calibration may only be required on an annual basis, but it is important to perform the
regular checks (e.g. on each use) to identify faults.
12.1
Replacement Parts
Item
Description
Part No
Suggested
Frequency of
Replacement
1
Analox MIR CO2 Bench 100mbar
MIRDJNX4S
5 years
2
Analox MIR CO2 Bench 20mbar
MIRAGNX4S
5 years
3
Analox MIR CO2 Bench 25mbar
MIREGNX4S
5 years
4
Analox MIR SS CO2 Bench 20mbar in
Helium
MIRASHX4S
5 years
2
Analox Hyperbaric/Hypobaric Oxygen
Sensor 0 to 600mbar pO2 (hypobaric), 0
to 1500 or 2000mbar pO2 (hyperbaric)
ranges
9100-9212-9HSUBA
2 years
3
Analox Hyperbaric Oxygen Sensor 0 to
3000mbar pO2 ranges
9100-9212-9H3
2 years
4
Analox REM2 Replacement Temperature/
Humidity Sensor
DD1TUPHAT02
5 years
5
Immersible breather Vent
SA2 XMRCBL/D6693
5 years
6
MkIII Operator Console LCD display
module
*
2553-9200S
5 years
7
Replacement battery
–
Operator Console
Real Time Clock (2 required per console)
9430-4794
5 years
*
The Console must be returned to Analox for replacement
12.2
Accessories
Item
Description
Part No
1
Calibration Gas
Contact Analox for exact requirements
2
Flow or pressure regulators and flow meters
Contact Analox for exact requirements
3
Humidity Calibrator 0% RH and 75%
SA1P16RH
4
Push in flow adaptor
8000-0069A
5
Calibration tubing
–
neoprene (Qty of 1m)
1817-5000