Ax60+ Multi-Gas
Service Manual
Int. Approved
Document ref: P0159-803-11
September 2019
Page 52 of 62
Copyright © 2015 Analox Ltd. All Rights Reserved.
Commercial in Confidence
10
Ax60+ Kiosk Option
The Ax60+ Kiosk can be serviced in the same way as the standard Ax60+ by temporarily fitting a portable calibration unit
(part no. AX60CNQNXA). This is supplied with a pre-fitted cable terminating in an RJ45 plug, an RJ45 cable coupler and a
length of cable:
AX60CNQNXA portable calibration unit
AX60CNQXA temporary cable and coupler
To connect the AX60CNQNXA calibration unit to the Ax60+ Kiosk Sensor, follow this procedure:
WARNING: DISCONNECT AND ISOLATE THE AX60+ KIOSK FROM THE MAINS POWER SUPPLY
BEFORE OPENING THE CO
2
SENSOR ENCLOSURE.
[1]
Remove the front cover from the Ax60+ Kiosk Sensor enclosure.
[2]
Install the temporary cable through the cable gland and connect as follows:
ORG PAIR (temporary cable); Black (existing cable)
BRN PAIR (temporary cable); Black (existing cable)
GRN (temporary cable)
GRN/WHT (temporary cable)
ORG PAIR (existing cable)
BRN PAIR (existing cable)
GRN/WHT (existing cable)
GRN (existing cable)
BLU/WHT (existing cable)
SPARE (not used)
[3]
Replace the front cover on the Ax60+ Kiosk Sensor enclosure.
[4]
Reconnect the mains supply and power-up the Ax60+ Kiosk.