ACG+
Service & Troubleshooting Manual
Int. Approved
Document ref: P0149-825-00
October 2022
Page 28 of 37
Copyright © 2022 Analox Ltd. All Rights Reserved.
Commercial in Confidence
4.3.1
Replacing the flow adaptor orings
1]
Remove the flow adaptor for the sensor and remove the old oring.
2]
Take a new 11.1mmOD x 8.1mmID oring (2323-0017X) and refit to the flow adaptor,
replace the flow adaptor to the sensor.
4.3.2
Replacing the PID cell
The following process describes how to replace the entire PID component if it is damaged or badly
contaminated. The replacement part number for your sensor’s PID is:
9100-2060S.
1]
To replace the PID component, remove the flow adaptor from the sensor inlet.