background image

 

ADL6012-EVALZ

 User Guide

UG-1675

One

 

Technology

 

Way

 

 

P.O.

 

Box

 

9106

 

 

Norwood,

 

MA

 

02062-9106,

 

U.S.A.

 

 

Tel:

 

781.329.4700

 

 

Fax:

 

781.461.3113

 

 

www.analog.com 

 

Evaluating the 

ADL6012

 2 GHz to 67 GHz, 500 MHz Bandwidth, Envelope Detector 

 

 
PLEASE SEE THE LAST PAGE FOR AN IMPORTANT  
WARNING AND LEGAL TERMS AND CONDITIONS.

 

Rev. 0 | Page 1 of 9 

FEATURES 

Fully featured evaluation board for the 

ADL6012

 

Specified up to 67 GHz 
3.15 V to 5.25 V operation 

EVALUATION KIT CONTENTS 

ADL6012-EVALZ evaluation board 

EQUIPMENT NEEDED  

Analog signal generator up to 67 GHz (E8257D) 
High speed oscilloscope (MSOS254A) 
DC voltmeter (34401A) 
5 V power supply (E3631A) 
Differential probe (1131A) 

GENERAL DESCRIPTION 

The ADL6012-EVALZ provides efficient evaluation of the 

ADL6012

 2 GHz to 67 GHz envelope detector. The 

ADL6012

 

has fast rise time and fall time responses, around 1 ns 
depending on input power levels, with an output envelope 
bandwidth of 500 MHz. The ADL6012-EVALZ provides the 
basic connections for the RF input signal and the envelope 
output voltage. Differential envelope outputs on the ADL6012-
EVALZ can be directly connected to a high speed oscilloscope 
to evaluate the differential envelope tracking capabilities. 

For full details on th

ADL6012

, see th

ADL6012

 data sheet. 

Consult th

ADL6012

 data sheet in conjunction with this user 

guide when using the ADL6012-EVALZ evaluation board.  

EVALUATION BOARD PHOTOGRAPH 

220

21-

001

 

Figure 1.  

Содержание ADL6012-EVALZ

Страница 1: ...power supply E3631A Differential probe 1131A GENERAL DESCRIPTION The ADL6012 EVALZ provides efficient evaluation of the ADL6012 2 GHz to 67 GHz envelope detector The ADL6012 has fast rise time and fal...

Страница 2: ...Evaluation Board Setup Equipment 3 Power Supply 3 RF Input 3 High Speed Oscilloscope 3 DC Voltmeter 3 Evaluation Board Setup Procedure 4 DC Coupled Constant Envelope Output Measurements 4 High Speed P...

Страница 3: ...y measure the rise and fall times of the ADL6012 The Keysight MSOS254A oscilloscope must have a bandwidth of 2 GHz or higher to accurately measure the ADL6012 output response The amplitude modulation...

Страница 4: ...age across the VEH and VEH turrets The Agilent 34401A voltmeter is used to take the dc output measurements as the RF input power varies across the ADL6012 detection range HIGH SPEED PULSED OUTPUT MEAS...

Страница 5: ...TIAL PROBE E8257D SIGNAL GENERATOR 22021 003 POWER SUPPLY E3631A 5V Figure 3 ADL6012 Envelope Measurement Using Differential Probe 50 50 22021 004 KEYSIGHT MS0S254A POWER SUPPLY E3631A 5V E8257D SIGNA...

Страница 6: ...L 69157 102HLF ELF40 002 142 0701 851 DUT C7 C8 C5 C2 CAL2 CAL1 RFIN R12 R11 R9 R10 R4 C6 R3 DIS VEH _TP VEH _TP VEH VEH J2 J1 ACOM_VOCM R6 R5 R7 GND1 C3 VPOS GND2 R2 C1 VOCM_B2 CAL_TRACE VEH _TP VEH...

Страница 7: ...RD THICKNESS 0 062 10 8mil ROGERS 4003 PRIMARY SILKSCREEN PRIMARY SOLDER MASK PRIMARY SIDE LAYER 1 GROUND PLANE LAYER 2 GROUND PLANE LAYER 3 GROUND PLANE LAYER 4 SECONDARY SOLDER MASK SECONDARY SILKSC...

Страница 8: ...Panasonic ERJ2RKF2002X 1 R4 Differential output load resistor open not installed 3 R5 to R7 Resistor divider network for output common mode voltage open not installed 2 R9 R10 Series envelope output r...

Страница 9: ...any reason Upon discontinuation of use of the Evaluation Board or termination of this Agreement Customer agrees to promptly return theEvaluationBoardto ADI ADDITIONALRESTRICTIONS Customermay not disas...

Отзывы: