Anaheim Scientific P771 Скачать руководство пользователя страница 1

 
 
 

 

User Manual 

 
Model P773 

Total Dissolved Solids (TDS) Meter 

 
 
 
 
 
 
 
 
 

Содержание P771

Страница 1: ...User Manual Model P773 Total Dissolved Solids TDS Meter ...

Страница 2: ...NFORMATION 7 9 1 SERVICE REPAIR OR CALIBRATIONS 7 9 2 WARRANTY 8 1 Introduction The P773 is a wide range pocket sized total dissolved solids meter perfect for use in field environments such as Hydroponics Fish farming Agriculture Swimming pools and spas Waste water systems Plating industry Food industry Pharmaceutical and Medical Cooling towers Chemical industry Having a rugged waterproof housing ...

Страница 3: ...thin the limits indicated in the Technical Specifications section of this manual Avoid making measurements in the presence of explosive gas combustible gas steam or excessive dust Do not install substitute parts or perform any unauthorized modifications to this instrument Return the instrument to Anaheim Scientific for service and repair to ensure that safety features are maintained The following ...

Страница 4: ...n initial physical inspection should be made to ensure that no damage has been sustained during shipment Inspect the packing box on receipt for any external damage If any external damage is evident remove the instrument and visually inspect its case and parts for any damage If damage to the instrument is evident a description of the damage should be noted on the carrier s receipt and signed by the...

Страница 5: ...ot touch or wipe the surface of the inner black sensor of the electrode 2 Open the battery compartment to take out the screwdriver 3 Press the button to turn the meter power on 6 2 Calibration 1 Dip the electrode into the calibration solution Stir gently and wait until the display is stabilized Adjust the reading to 94 940 ppm at 25 by turning the trimmer Span located at left side of battery compa...

Страница 6: ...trode 6 4 Changing the battery 1 The installed Lithium CR2032 batteries have a typical life span of 100 hours Replace the battery when the display begins to fade or flash 2 Loosen the battery compartment cap 3 Replace the old batteries with fresh Lithium CR2032 batteries noting the polarity 4 Replace the battery compartment cap tightly 5 Recalibrate the meter after changing the batteries 6 5 Elect...

Страница 7: ...age Environment 0 C to 50 C 0 to 90 R H Power Two 3 V Lithium CR2032 Batteries Expected Battery Life 100 hours continuous Display 3 Digit 21 x 18 mm LCD 8 x 7 Dimensions 33 5 mm x 170 mm 1 3 x 6 7 Calibration 1 point manual via trimmer span Weight 85 g 3 oz Housing ABS IP57 To ensure the most current version of this manual please download the current version here http anaheimscientific com images ...

Страница 8: ...repair and or calibration with your product s model number Once you receive a reply from service anaheimscientific com you will be asked to ship prepaid to the address below Package the unit carefully using filler or bubble wrap and if possible ship in original box Ship each unit separately Anaheim Scientific is not responsible for any shipping damage that may occur Include a packing list with eac...

Страница 9: ...ales receipt To obtain warranty coverage in the U S A this product must be registered by completing a warranty registration form on anaheimscientific com within fifteen 15 days of purchase Exclusions This warranty does not apply in the event of misuse or abuse of the product or as a result of unauthorized alterations or repairs The warranty is void if the serial number is altered defaced or remove...

Страница 10: ... Page intentionally left blank ...

Страница 11: ... Page intentionally left blank ...

Страница 12: ...22820 Savi Ranch Parkway Yorba Linda CA 92887 www anaheimscientific com 2012 Anaheim Scientific ...