Anaheim Scientific M150 Скачать руководство пользователя страница 1

 

 

 

Model M15

0

 

Mini 

UV

-

AB Meter

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание M150

Страница 1: ...Model M150 Mini UV AB Meter ...

Страница 2: ...M150 Features 3 2 SAFETY SUMMARY 4 3 COMPLIANCE STATEMENTS 5 4 PRODUCT CONTENTS AND INSPECTION 6 5 DEVICE DESCRIPTION 7 5 1 LCD Display 7 5 2 Control Buttons 7 6 DEVICE OPERATION 8 6 1 Power Button 8 6 2 Zero Button 8 6 3 Backlight Display 8 6 4 Unit Button 9 6 5 Data Hold Button 9 6 6 Battery Replacement 10 7 TECHNICAL SPECIFICATIONS 11 8 SERVICE REPAIRS CALIBRATION 12 9 LIMITED TWO YEAR WARRANTY...

Страница 3: ...o use use the M150 for 1 1 M150 Applications Monitoring blue light radiation hazards in welding UV sterilization Graphic arts Photochemical matching Photo resist exposure Curing of inks adhesives and coatings 1 2 Laboratory Applications Weathering degradation studies UV sterilization Virology Microbial genetics DNA research Biologic hoods General laboratory use 1 3 The M150 Features Professional h...

Страница 4: ...ual on off by push button or auto shut off after 15 minutes 2 Safety Summary CAUTION Adhere to the following conditions for safe and effective usage of this meter This is a precision device During use or storage do not go beyond its specifications to prevent any possible damage or danger Do not operate the meter in an environment filled with combustible gas liquids dust or fibers In order to avoid...

Страница 5: ...ent has been designed for usage within an environment of Pollution Degree 2 Clean the device with a dry soft cloth Wet cloths liquid and water are prohibited Operating Environment 5 C 40 C 80 RH 2km altitude Storage Environment 10 to 60 C 70 RH 3 Compliance Statements Caution This symbol indicates that the equipment and its accessories are subject to special collection and disposal procedures This...

Страница 6: ...emove the instrument and visually inspect its case and parts for any damage If damage to the instrument is evident a description of the damage should be noted on the carrier s receipt and signed by the driver or carrier agent Save all shipping packaging for inspection Forward a report of any damage to the agent through which the unit is procured Retain the original packing in case subsequent repac...

Страница 7: ...2013 20130412 5 Device Description 5 1 LCD Display 1 Main display 2 Value hold 3 ZERO display 4 Low power indicator 5 µW cm 2 unit display 6 mW cm 2 unit display 7 Decimal point display 5 2 Control Buttons 1 Power switch button 2 Backlight button 3 Hold lock button 4 Zero button 5 Unit button ...

Страница 8: ...put the device into sleep mode The display changes from light to dark 6 2 Zero Button Press the button for the zero adjustment if any digits are appearing when the sensor is covered If performing the zero adjustment after powering on several digits may not disappear In this case perform the zero adjustment again 6 3 Backlight Display Press key to turn backlight on Press the button again to power o...

Страница 9: ...er and put the device to operating mode The screen displays 0000µW cm 2 Press the button to switch from 0000µW cm 2 to 00 00mW cm 2 To select a different unit just press this button once again 6 5 Data Hold Button Press the button to go into hold mode and HOLD appears on the screen to allow you to read the data Press this button once again to deactivate it ...

Страница 10: ...o change without notice Anaheim Scientific 2013 20130412 6 6 Battery Replacement WARNING If the symbol appears on the LCD please replace the battery immediately Remove the battery cover Noting the polarity replace the battery Install the battery cover ...

Страница 11: ... temperature induced error 0 38W Cm 2 from 25 Range 3999 µW cm 2 39 99mW cm 2 Accuracy Attenuation 3 year UV Spectrum 290 nm to 390 nm Peak Sensitivity Wavelegth of 365 nm Display 3 digits LCD with max reading 4000 Units µW cm 2 and mW cm 2 Sampling Rate About 3 seconds Power Supply 2 AAA batteries Battery life 50 hours during consecutive use Dimensions 133 48 27 mm H W D Weight 90g Warranty Limit...

Страница 12: ...naheimscientific com you will be asked to ship prepaid to the address below Package the unit carefully using filler or bubble wrap and if possible ship in the original box Ship each unit separately Anaheim Scientific is not responsible for any shipping damage that may occur Include a packing list with each unit shipped stating what type of service is required and include the return shipping inform...

Страница 13: ...130412 shipping Return shipping to locations in North American is included for Warranty Service For overnight shipments and non North American shipping fees please contact Anaheim Scientific Anaheim Scientific ATTN Service Repair 22820 Savi Ranch Parkway Yorba Linda CA 92887 www anaheimscientific com ...

Страница 14: ...les receipt To help us better serve you please complete the warranty registration for your new instrument via our website www anaheimscientific com Exclusions This warranty does not apply in the event of misuse or abuse of the product or as a result of unauthorized alterations or repairs The warranty is void if the serial number is altered defaced or removed Anaheim Scientific shall not be liable ...

Страница 15: ...15 anaheimscientific com Technical data subject to change without notice Anaheim Scientific 2013 20130412 ...

Страница 16: ...16 anaheimscientific com Technical data subject to change without notice Anaheim Scientific 2013 20130412 22820 Savi Ranch Parkway Yorba Linda CA 92887 www anaheimscientific com 2013 Anaheim Scientific ...