11
3.1
Overview of both special boot-up modes
There are two special boot-up modes on the e-bx. These modes are very useful tools for technicians to determine
the cause or sources of power option failure; to boot-up actuators that may have lost their positioning; or to
replace a defective actuator related to the e-bx. Each mode has its own distinctive features.
3.2
Activating special boot-up modes
When the button underneath the e-bx is held during system start-up, the green LED flashes with a delay of two
seconds. The number of flashes corresponds to the mode required. Release the button when the number of
flashes equals the mode required. Therefore, if the first mode is required, release the button at the first green LED
flash and if the Second mode is required, release at the second flash.
Tables in this section provide a summary of features for each mode.
3.3
First special boot-up mode
This mode was designed to help determine if an actuator is defective or if it stopped moving because of a problem
with the controller. This mode can also be used for an actuator position reset. In this mode, all movement
limitations will be disabled except for minimum and maximum limits for each actuators under the "General"
submenu (section 2.1.2.1.)
Table for the first special boot-up mode
Number of green LED
flashes at start-up on
the e-bx
Mode 1 indications
What becomes
disabled
Applicable actuators
Features
1
•
Red light left
of the
pushbuttons
(SWKEYPAD)
flashes
constantly.
•
Power options
are displayed
in “seating”
on
joystick screen
Actuator movement
limits are disabled. If
actuator positioning
system is defective,
movement will not be
possible.
•
Tilt (1). See
Note 1
•
Recline (2). See
Note 1
•
Seat elevate (3).
See
Note 1
•
Center mount legrest
(4). See
Note 2
•
Left legrest (5).
See
Note 1
•
Right legrest (6).
See
Note 1
•
Left stander (5).
See
Note 2
•
Right stander (6)
.
See
Note 2
•
Allows a minimum and maximum
position reset of any actuator that
are not defective.
•
Works with joystick and pushbutton
keypad (SWKEYPAD)
•
Possible error warnings are not
disabled
•
The stander operates normally and
will automatically soft stop at the
comfortable position.
•
The chair is not restricted in its
driving by default except for normal
limit settings.