OPERATING AND MAINTENANCE MANUAL

- CONTINUOUS / INTERMITTENT -

DIGITAL & ANALOG VACUUM REGULATOR 

®

1111 Lakeside Drive, Gurnee, IL 60031-4099 | Tel: 1.800.448.0770 | Fax: 847.855.6300 

www.ohiomedical.com

25B East Pearce St., Richmond Hill, ON L4B 2M9 Canada | Tel: 

905.764.7736

 | Fax: 905.764.7743 | 

www.amvex.com

0413

0413

Models Included:

DIGITAL (D)

ANALOG (A)

VR-CIU2-F2D

VR-CIU2-F2A

VR-I2U2-F2D

VR-I2U2-F2A

VR-PIU2-F2D

VR-PIU2-F2A

VR-PPU2-F2D

VR-PPU2-F2A

VR-NIU2-F2D

VR-NIU2-F2A

VR-NNU2-F2D

VR-NNU2-F2A

Rx ONLY

Содержание VR-CIU2-F2D

Страница 1: ... 448 0770 Fax 847 855 6300 www ohiomedical com 25B East Pearce St Richmond Hill ON L4B 2M9 Canada Tel 905 764 7736 Fax 905 764 7743 www amvex com 0413 0413 Models Included DIGITAL D ANALOG A VR CIU2 F2D VR CIU2 F2A VR I2U2 F2D VR I2U2 F2A VR PIU2 F2D VR PIU2 F2A VR PPU2 F2D VR PPU2 F2A VR NIU2 F2D VR NIU2 F2A VR NNU2 F2D VR NNU2 F2A Rx ONLY ...

Страница 2: ... Cover Screws 4 pcs 2 Mode Face Assembly 3 Mode Face Assembly 3 Mode 2 Mode Intermittent 17 Back Plate Screws 2 pcs 18 Bottom Plate Screw 19 Orifice 20 Regulating Module 23 Pediatric Relief Screw 24 Pediatric Relief Body 25 Pediatric Relief Ball 26 Pediatric Relief O ring 27 Intermittent Module 28 Chamber O ring 0 200mmHg Gauge Analog or Digital 0 300mmHg Gauge Analog or Digital 0 160mmHg Gauge An...

Страница 3: ... to room temperature before use User Responsibility This device is to be used ONLY by people who have been properly trained on the operation of the device Operation of this device is not to be done if flammable anesthetics are present due to the possibility of explosion caused by static charge This product performs as explained in this manual as long as the assembly use repair and maintenance are ...

Страница 4: ...tubing is recommended by NFPA National Fire Protection Association NFPA 99 2002 Healthcare facilities pages 497 498 To prevent possible contamination of the regulator a high flow suction filter or an overflow safety trap provided by Amvex is recommended between the regulator and the collection canister NFPA recommends use of an overflow trap to protect the vacuum regulator outlet and vacuum system...

Страница 5: ...slowly to reach various vacuum rates Ensure that the level of vacuum maintains consistently when the vacuum tubing is kinked 5 Move the selector switch to INT 6 Kink vacuum tubing 7 Timing cycles are approximately 16 seconds on and 8 seconds off NOTE The intermittent unit starts in the off cycle 8 Decrease the vacuum level to zero and move the selector switch to the OFF position Pediatric Neonatal...

Страница 6: ...body during timing NOTE The off time must be adjusted prior to the on time 6 The unit will begin in the off mode of the intermittent cycle To increase the off time turn the OFF Needle Valve clockwise To decrease the off time turn the OFF Needle valve counterclockwise 7 After the off time has been adjusted to the desired timing the on time may be adjusted To increase the on time turn the ON Needle ...

Страница 7: ... mixture may cause small surface cracks to some of the plastic parts that may not be apparent to the user CAUTION Do not steam autoclave immerse in liquid or gas sterilize the Vacuum Regulators This may damage the unit CAUTION If Vacuum Regulator becomes contaminated internally warranty is voided Do not send Vacuum Regulator back to the manufacturer Follow your facilities procedures for handling c...

Страница 8: ...operated under conditions of normal use regular periodic maintenance and service is performed and repairs are made in accordance with this operation manual The warranty period for all expendable parts of the Product is sixty 60 days from the date the Company ships the Product to the customer The foregoing warranty shall not apply if the Product has been repaired or altered by anyone other than the...