Amti Gen5 Скачать руководство пользователя страница 5

 

012216 

Page 5 of 6 

ver1 

 

e.

 

If the lists do not automatically populate check your connections and that the amplifier has 

power and click 

‘Find Amplifiers’

. If you are still having trouble consult the Gen 5 manual. 

f.

 

Under ‘

Global Settings’

, and 

‘Digital Outputs’

, select English or metric units. 

g.

 

Under ‘

Global Settings’

, and 

‘Analog Outputs’

 select either MSA 6 Compatible or Fully 

Conditioned.  

 

If you are using the Gen 5’s digital output only (for example if you are collecting through 

NetForce, or integrating digitally with a third party system) this setting doesn’t matter. 

Leave it as ‘

MSA 6 Compatible

’ 

 

If you plan to use the analog output from the Gen 5, choose 

‘Fully Conditioned

’ 

 

If you are unsure leave it as ‘

MSA 6 Compatible

’. You can always change this setting 

later. Consult the Technical note 

Gen 5 Setup and Optimization 

for more details on 

what these modes represent.  

h.

 

Click on 

‘Apply’

 and ‘

Save’

 

i.

 

Click ‘

Done’

 

j.

 

In the next window select the location where data collected through NetForce should be saved. 

Navigate to an appropriate folder or select the folder 

‘Data’ 

under 

‘AMTINetForce’ 

and click 

‘Open’

, then ‘

Open’

.  

k.

 

Click 

‘OK’

 for the location of the default directory.  

l.

 

Click ‘

OK

’ to complete the setup and close the program. 

 

You can find more detailed information about setting up and running NetForce in the 

NetForce User 

Manual. 

 

 

6.

 

Gen 5 Amplifier Configuration 

 

In order to properly configure your force plate you must select 

Excitation Voltages

 and A

mplifier Gains

 

that are relevant to the measurements you intend to make. You can choose different values for each 

channel. If you plate to use the analog output you will also need to set the 

Analog Sensitivities

. For 

information on how to do this consult the Technical note 

Gen 5 Setup and Optimization

 

 

Excitation Voltage: 

To achieve the best signal to noise ratio you want the Gen 5 to output the largest possible excitation 

voltage. This will normally be 10 volts. 

 

Amplifier Gain: 

To achieve the best system resolution, apply the largest gain setting available where the Gen 5 

measurement range is greater than the expected application range. 

 

Typical Settings: 

NOTE: 

The settings in the table below are simply guidelines and may not be suitable for your particular 

study set. These can be used as a starting point but you should verify their appropriateness under your 

conditions and ensure that none of the channels are saturating.  

 

 

 

 

Содержание Gen5

Страница 1: ... you plan to use a third party system for data collection Step 5 Open Netforce and run through the setup helper instructions on Page 4 of this guide Step 6 Configure the system with settings relevant to your research For common settings see page 4 of this guide More detailed information on the process can be found in the technical note Gen 5 Setup and Optimization found on your USB drive Step 7 Vi...

Страница 2: ...nents Force Plate or transducer Force transducer to amplifier cable 9615 cable Amplifier Gen 5 Power supply USB cable Connections a Connect the 9615 cable from the force transducer to the amplifier b With the power switch set to off plug the power supply into the amplifier then into a socket c Plug the USB cable from the amplifier to the computer if you have more than one amplifier plug them all i...

Страница 3: ...n its logo d Right click the GEN 5 device and select Update Driver Software e Select Browse my computer for driver software f Click Browse and navigate to the folder containing the driver files Make sure to select the appropriate 32 bit or 64 bit version for your operating system and then click OK i To determine which operating system you have click Computer under the Windows Start menu then right...

Страница 4: ...g NetForce 5 NetForce Setup a The first time you open NetForce it will automatically take you through an initial setup If this does not open automatically access it by clicking Setup then Setup Helper and Yes b The first window will ask if you wish to continue with the setup click Yes c Next select the configuration that matches your hardware in this case highlight Optima Gen 5 using USB 2 0 and c...

Страница 5: ...ick Open then Open k Click OK for the location of the default directory l Click OK to complete the setup and close the program You can find more detailed information about setting up and running NetForce in the NetForce User Manual 6 Gen 5 Amplifier Configuration In order to properly configure your force plate you must select Excitation Voltages and Amplifier Gains that are relevant to the measure...

Страница 6: ...Step 5 b At the bottom left of each graph set the Amp ID to the force plate number for which you wish to view the data This should be 1 or greater no data will appear if it is set to 0 c Under the Startup menu select Hardware Zero The Auto Zero light on the amplifier should flash d Under the Settings menu select Acquisition Settings and ensure that the Mode at the bottom right of the window is set...

Отзывы: