No Calibration Pen Type pH Tester AMT28   

USER GUIDE 

Quick Step for Use:   

1.

 

First  time  use  or  long  time  not  use,  turn  off  the meter  and  put 
sensor  in  the  sample  liquid  or  tap  water 

(don’t  put  sensor  in 

distilled water) for about 2-5 minutes. Shake it gently in the water.   

2.

 

Long press ON/OFF button to start test, meter will display pH and 
temperature  value  alternately.  Value  will  automatic  hold  when 
stable.   

3.

 

Short press ON/OFF, meter will test again.   

4.

 

Long press ON/OFF meter will shut off.   

5.

 

After 5 months or accuracy not good,    meter in shut down state, 
long  press  about  15  seconds,  till  meter  display 

“dEFA",  release 

button and meter will reset to factory default.   

 

Calibration (No need, this function only for  professional user  or 
meter work wrong!) 

1.

 

Though this meter no need calibration and easy to reset to factory 
default, it also can calibration by professional user.   

2.

 

Shut down meter, put sensor in the 6.86 buffer solution, shake it 
gently and wait for 2 minutes. 

3.

 

Long  press  ON/OFF  about  10  seconds  till  meter  display  “CALI”, 

meter will automatic calibration and no need any operation. After 
calibration finish, meter will turn off.   

 

Maintenance:   

1.

 

Always clean the sensor after use by tap water.   

2.

 

Check  the  battery  compartment  to  avoid  water  into  or  battery 
leakage.   

 