7.3.5.5 Restore to Factory Setting
Touch "Restore Factory Settings" and enter the password to restore the screen
brightness, alarm volume, system date and time, accuracy factor and other
parameters to the system factory default values. System log records will not be
deleted after factory setting restore. The follows items will be restored:
NO
Item
Restore to
1
Feeding Parameters
0
2
RFID
Switch On
3
Brightness
4
4
Alarm volume
1
5
Auto-Lock Time
Off
6
1)
Night Mode
Switch On
2)
Start time
7:00PM
3)
End time
7:00AM
4)
Brightness
1
5)
Alarm volume
1
7
KTO (KTO between meals)
Switch on
Caution
Please be cautious to restore factory settings.
8. Alarm
Overview
The
AMSure
® Enteral Feeding Pump has alarms that are broken into three
different priorities: High Priority, Medium Priority, and Low Priority. Light
signals are listed as below. Alarm signal can be perceived correctly at distance
of 4m from the alarm system. The delay time of triggering the alarm signal is
not more than 2s.
•
High Priority (Red Blinking Indicator Light),
•
Medium Priority (Yellow Blinking Indicator Light),
•
Low Priority (Solid Yellow Indicator Light)
These alarms occur based upon feedback from different sensor inputs from the
pump. Key inputs for the alarms include the following:
•
Motor Optocoupler Voltage
•
Battery Voltage
•
Upstream Infrared Sensor Voltage