“Power” button
and touch the “Switch Off” button on the screen to turn
off the pump. User also can hold the “Power” button
for 4 seconds to
turn off the pump directly.
Fig 7.2.7-2 Switch Off
Caution
Close the roller clamp, disconnect the tubing from the patient’s enteral
access before unloading tube from the pump. Otherwise, the free flow may
occur by gravity causing over feeding and harm to patient if there is still
nutrition in the container.
7.2.8 KTO
The Keep Tube Open (KTO) function within the pump will help prevent a
blockage of the feeding set by moving the feeding solution periodically
within the tube. On regular intervals, calculated from the pump’s presently
programmed feed rate, a small and clinically insignificant portion of feeding
solution (0.2mL/bolus) is moved through the tube.
KTO includes Automatic KTO for intermittent feeding mode only and Manual
KTO when pump is running for both continuous mode and intermittent mode.
Press the power button to
turn off the pump
Hold the power button to turn
off the pump