7.1.3.2 Self-Testing
After the pump is switched on, the pump will automatically perform a safety
self-check. After the self-check is passed, user will hear two short beeps of
“DiDi”. The system automatically enters to the next step. If the self-test fails,
the system will display the following prompt message
“System error detected.
Please contact your customer service.
Fig 7.1.3.2-1 Self-Testing Failed
Caution
When self-testing failed, pump is not allowed to run for feeding, but user still
can check other interfaces through Home button.
7.1.4 Feeding Set Identification System
To guarantee the feeding accuracy of the pump, only the feeding sets listed in
chapter 9
“Dedicated Enteral Feeding Sets (Disposable)” are recommended.
RFID identification technology when is enabled on the pump is one of the
solutions for user to avoid using incorrect sets.
The pump confirms whether it is a dedicated enteral delivery set by swiping the
RFID tag on the feeding set.