12-17
Troubleshooting
Operator Manual
129373-496
Too Long In Jacket
Charge (Continued)
Screen with Operator Instructions
Screen with Service Instructions
Occurs if door seal
does not reach
5 psig (0.34 Pbar) within
allotted time.
12.4.3 Too Long To Seal
Door
Alarm
Description
Screen with Service Instructions
STATUS..
SERVICE INFORMATION:
232
TOO LONG IN JACKET CHARGE
→
→
→
→
→
JACKET DID NOT REACH REQUIRED
TEMPERATURE WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
1.
STEAM PRESSURE LESS THAN 50 PSIG
→
→
→
→
→
CHECK STEAM SUPPLY PIPING
2.
STEAM REGULATOR MALFUNCTION
→
→
→
→
→
REPAIR
3.
SOLENOID VALVE MALFUNCTION
→
→
→
→
→
REPAIR S09
4.
CONTROL OUT OF CALIBRATION
→
→
→
→
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
PAPER
SERVICE
EXIT
FEED
MODE
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
227
TOO LONG TO SEAL DOOR
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
REMAIN IN ALARM CONDITION UNTIL
DOOR IS SEALED
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK STEAM SUPPLY VALVE
→
→
→
→
→
IF CLOSED, OPEN VALVE AND
RESELECT CYCLE
3. IF ALARM RECURS, CALL SERVICE
STATUS..
SERVICE INFORMATION:
228
TOO LONG TO SEAL DOOR
→
→
→
→
→
DOOR SEAL DID NOT REACH 5 psig
WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
1.
SEAL PRESSURE SWITCH MALFUNCTION
→
→
→
→
→
READJUST PS1(PS2)
→
→
→
→
→
REPAIR PS1(PS2)
2.
SEAL NOT ACTIVATING
→
→
→
→
→
CHECK SEAL STEAM
→
→
→
→
→
CHECK SEAL EXHAUST
3.
SOLENOID VALVE MALFUNCTION
→
→
→
→
→
REPAIR S35(S36)
PAPER
CLEAR
EXIT
FEED
ALARM