10-32
129373-496
Operating Instructions
Out of Cycle Options
10.10.6 RS-232 Interface
This function allows supervisor to download cycle data into the device (either
a computer or printer) connected at the RS-232 interface port.
1. Press
RS-232 INTERFACE
on screen #51.
51
SUPERVISORY CONTROL
RETURN
MAIN
MENU
CHANGE
CHANGE
CUSTOMER NAME
ACCESS CODES
CHANGE
DISPLAY
STERILIZER ID
ACCESS CODES
DEFAULT
RS-232
VALUES
INTERFACE
Pressing RETURN returns display to
screen #87.
Pressing MAIN MENU returns display to
screen #1.
2. Screen #60 allows supervisor to indicate the device which will receive the
downloaded cycle data. Select the correct device by pressing the appro-
priate button. Current device setting appears under the screen title.
Display automatically returns to screen #51 once a device is selected.
60
RS-232 INTERFACE
TO COMPUTER
TO COMPUTER
TO PRINTER
CANCEL
Pressing CANCEL returns display to
screen #51.
•
To set up computer interface, plug a null modem RS-232 cable from COM1
of the sterilizer control to COM1 or COM2 of a personal computer. The
data may be retrieved from the computer by use of a standard commu-
nication software, or by use of the STERIS RS-232 software program. The
STERIS RS-232 software program will retrieve the data and store it in a
pre-defined file.
•
To set up printer interface, plug a null modem RS-232 cable from COM1
of the sterilizer control to the printer.