![Amsco Century Скачать руководство пользователя страница 113](http://html1.mh-extra.com/html/amsco/century/century_operators-manual_2936190113.webp)
10-2
129373-496
Operating Instructions
Out of Cycle Options
Screen #87 lists the remaining out-of-cycle functions. To access any of the
listed functions, refer to the description, included in this section, titled the same
as the button on screen #87.
To exit the option menu screens, press
MAIN MENU
on screen #13 or #87.
Display returns to the main menu (screen #1).
Status Print is used, when the sterilizer is out of cycle, to generate a printed
record of the current chamber status.
Press
STATUS PRINT
on screen #13 to automatically generate a printout
listing the current temperature and pressure in the sterilizer chamber at the
time the button was pressed (see Figure 10-1).
10.2 Status Print
87
OUT OF CYCLE OPTIONS
PREVIOUS
SUPER-
VISORY
DISPLAY
SERVICE
SENSORS
MODE
STANDBY
MAIN
MENU
Pressing PREVIOUS returns display to
screen #13.
Pressing MAIN MENU returns display to
screen #1.
13
OUT OF CYCLE OPTIONS
NEXT
MAIN
MENU
STATUS
CHANGE
VALUES
DUPLICATE
DISPLAY
VALUES
PAPER
FEED
VALUES
Pressing NEXT advances display to
screen #87.
Pressing MAIN MENU returns display to
screen #1.
STATUS PRINT
------------------------
TIME OF DAY
1:53:30P
CHAMBER PRESS
0.6 pisg
CHAMBER TEMP
78.3C
JACKET TEMP
115.0C
WASTE TEMP
25.0C
Figure 10-1. Example of
Out-of-cycle Status Printout