![Amsco C Series Скачать руководство пользователя страница 78](http://html1.mh-extra.com/html/amsco/c-series/c-series_operators-manual_2936189078.webp)
7-2
P129394-176
Operator Manual
Cycle and Control Value Programming
7.2 Change Cycle
Values
Refer to Figures 7-1 through 7-3.
7.2.1 Overview
1.
Press the
ON
touch screen pad, if the sterilizer is in STANDBY.
The control advances to Status screen #1.
2.
At screen #1, press the
MENU
touch screen pad, and the control
advances to screen #2; access the cycle values by pressing
CHANGE CYCLE VALUES
.
3.
Pressing
CHANGE CYCLE VALUES
advances the control to
screen #10, prompting the operator to select a cycle to change.
4.
Press one of the displayed cycles, or
MORE CYCLES
to find a
cycle not currently on the display. Once the cycle has been
found and selected, the control advances to screen #11,
showing basic cycle information.
5.
Select the value to change:
NAME
(name of the cycle),
TEMP
(sterilize exposure temperature),
STER
(sterilize exposure time),
or
DRY
(drying phase time). Screen #11 can also be used to
enter up to two
CUSTOM
cycles names.
6.
Press the touch screen touchpad; the display advances to a
screen for making these changes.
7.
Once all selections have been made, press the
EXIT
touch
screen pad.
7.2.2 Adjusting Cycle
Sterilize Time and Dry Time
The following selections can be adjusted:
•
NAME:
For changing the cycle type.
•
Prevacuum Configuration –
PREVAC, GRAVITY
or
LIQUID
.
•
SFPP Configuration –
PREVAC, GRAVITY, SFPP
or
LIQUID
.
•
CUSTOM1
and
CUSTOM2
buttons on screen #11 can be
used to enter user-defined cycle names.
NOTE:
When selecting a name, all other cycle values default to
the factory set values for that cycle.
LIQUID
SFPP
GRAVITY
PREVAC
SELECT NEW CYCLE NAME
OR SELECT ANOTHER ITEM TO CHANGE
12
TEMP
STER
DRY
NAME
NONE
EXIT
Содержание C Series
Страница 62: ...6 18 P129394 176 Operator Manual Sterilizer Operation Figure 6 9 Typical Printout of a Liquid Cycle ...
Страница 146: ...10 2 P129394 176 Operator Manual Service Procedures Figure 10 1 Serviceable Components ...
Страница 153: ...10 9 Service Procedures Operator Manual P129394 176 Figure 10 1 Piping Schematic ...