9-10
P129394-176
Operator Manual
Troubleshooting
Screen with Operator Instructions
9.2.12 Generator
Drain Temperature
Probe Failure Alarm
Screen with Service Instructions
STATUS . . ALARM! GENERATOR DRAIN
233
TEMPERATURE PROBE FAILURE
CHAMBER: 70.0 F 0.0 psig
GEN DRAIN: 300.0 RF
STERILIZER WILL:
•
PREVENT NEW CYCLE FROM BEING
STARTED UNTIL ALARM IS CLEARED
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
234
GENERATOR DRAIN TEMP PROBE FAILURE
-> CHAMBER EXHAUSTED SLOWER THAN
THE EXPECTED RATE
CAUSES AND CORRECTIONS:
1. LOOSE CONNECTION IN PROBE WIRING
-> REPLACE
-> RECALIBRATE
2. PROBE FAILED
-> REPAIR S03
3. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4. MAIN CONTROL FAILURE
-> CHECK CONTROL BOARD POWER STATUS
LEDS
-> RECALIBRATE
PAPER
FEED
EXIT