4-2
P129394-068
Operator Manual
Techniques of Sterilization
4.2 Control Measures
for Verifying
Sterilization Process
A live spore test utilizing
B. stearothermophilus
is the most reliable
form of biological monitoring. This type of product utilizes controlled
populations of a controlled resistance, so that survival time and kill
time can be demonstrated.
To verify the process, insert the biological indicator in a test pack
and place pack on the bottom shelf. Run test pack through a typical
cycle. On completion, forward test pack and monitor to appropriate
personnel for evaluation. Refer to AAMI guidelines to conduct routine
biological monitoring.
4.2.1 Biological Monitors
Tests such as the Dart
®
(Daily Air Removal Test) or Bowie-Dick are
designed to document the removal of residual air from a sample
challenge load.
Run a Dart or Bowie-Dick test cycle daily before processing any
loads in a sterilizer equipped with prevacuum cycles. The first
prevacuum cycle of each day should be used to test the adequacy
of air removal from the chamber and load, so that steam can
penetrate the load. It is not a test for adequate exposure to heat in
terms of time-at-temperature.
Gravity Cycle
250°F
(121°C)
30 Minutes
30 Minutes Double-wrapped instrument trays.
Maximum weight per tray, 25 lbs
(11.3 kg)
ST-8:2008
Gravity Cycle
270°F
(132°C)
25 Minutes
15 Minutes Fabric Packs
ST-8:2008
Liquid
250°F
(121°C)
45
N/A
,
L
IQUID
C
YCLE
O
NLY
: N
UMBER
OF
1000
M
L
C
ONTAINERS
FOR
F
ULL
L
OAD
ST-8:2008
Leak Test
*
N/A
N/A
N/A
None
ST-8:2008
Dart Test
*
270°F
(132°C)
3-1/2
Minutes
1 Minute
Dart or Bowie-Dick Test Pack
ST-8:2008
Dart Warm-up
*
270°F
(132°C)
3 Minutes
1 Minute
None
N/A
*
Not adjustable.
Table 4-1. AMSCO 400 Series Prevacuum Medium Steam Sterilizer Cycles –
36H, 36SL, 48H, 48SL, 60H, 60SL
Cycles:
Sterilize
Temp.
Sterilize
Time
Dry
Time
Recommended Load
Refer to
Table 3 (page iii)
for
Recommended Quantities
Validation
Standard
Содержание 36 H
Страница 2: ......
Страница 14: ...xii P129394 068 Operator Manual Techniques of Sterilization LIST OF FIGURES Descrption Page ...
Страница 59: ...6 11 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 63: ...6 15 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 66: ...6 18 P129394 068 Operator Manual Sterilizer Operation Figure 6 9 Typical Printout of a Liquid Cycle ...
Страница 67: ...6 19 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 78: ...6 30 P129394 068 Operator Manual Sterilizer Operation Healthcare Capital Equipment ...
Страница 116: ...8 10 P129394 068 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 140: ...9 24 P129394 068 Operator Manual Troubleshooting Healthcare Capital Equipment ...
Страница 152: ...10 12 P129394 068 Operator Manual Service Procedures Healthcare Capital Equipment ...