4-1
Techniques of Sterilization
Operator Manual
P129394-068
4.1 General
The information in this section is intended as a general guide to
steam sterilization techniques. Also recommended is reference to
the standards of the Association for the Advancement of Medical
Instrumentation (AAMI ST-79, comprehensive guide to steam
sterilization and sterility assurance in health care facilities).
•
Prior to sterilization, all materials and articles must be thoroughly
cleaned.
•
After sterilization, goods should be stored in conditions that will
not compromise the barrier quality of their wrapping materials.
Important:
Applicable cycles have been validated to satisfy the
requirements outlined in
(page ii). If cycle parameters
(sterilize time, dry time, temperature) other than those listed are
required, it is the responsibility of the healthcare facility to consult
and follow the device manufacturer’s written instructions.
NOTE:
Contact STERIS Corporation for information on a wide range
of education/ training programs designed to meet the educational
needs of healthcare industries.
As part of the operator’s verification of the sterilization process,
biological indicators may be used to demonstrate that sterilization
conditions have been met.
NOTE:
Contact STERIS
®
for information on specific biological
indicators recommended for use with this sterilizer.
WARNING: It is inappropriate
for a healthcare facility to
sterilize liquids for direct
patient contact.
!
Cycle Default Status:
Factory-Default
Optional (Supervisor Access Only)
Table 4-1. AMSCO 400 Series Prevacuum Medium Steam Sterilizer Cycles –
36H, 36SL, 48H, 48SL, 60H, 60SL
Cycles:
Sterilize
Temp.
Sterilize
Time
Dry
Time
Recommended Load
Recommended Quantities
Validation
Standard
Prevacuum Cycle
270°F
(132°C)
4 Minutes
30 Minutes Double-wrapped instrument trays,
porous and non-porous loads.
Maximum weight per tray, 25 lbs
(11.3 kg)
ST-8:2008
Prevacuum Cycle
275°F
(135°C)
3 Minutes
30 Minutes
ST-8:2008
Prevacuum Cycle
270°F
(132°C)
4 Minutes
20 Minutes Fabric Packs.
ST-8:2008
Prevacuum Cycle
270°F
(132°C)
4 minutes
5 Minutes
Single Fabric Pack
ST-8:2008
Gravity Cycle
250°F
(121°C)
30 Minutes 15 Minutes Fabric Packs
ST-8:2008
Gravity Cycle
270°F
(132°C)
15 Minutes
30 Minutes Double-wrapped instrument trays.
Maximum weight per tray, 25 lbs
(11.3kg)
ST-8:2008
TECHNIQUES OF STERILIZATION
4
Содержание 36 H
Страница 2: ......
Страница 14: ...xii P129394 068 Operator Manual Techniques of Sterilization LIST OF FIGURES Descrption Page ...
Страница 59: ...6 11 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 63: ...6 15 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 66: ...6 18 P129394 068 Operator Manual Sterilizer Operation Figure 6 9 Typical Printout of a Liquid Cycle ...
Страница 67: ...6 19 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 78: ...6 30 P129394 068 Operator Manual Sterilizer Operation Healthcare Capital Equipment ...
Страница 116: ...8 10 P129394 068 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 140: ...9 24 P129394 068 Operator Manual Troubleshooting Healthcare Capital Equipment ...
Страница 152: ...10 12 P129394 068 Operator Manual Service Procedures Healthcare Capital Equipment ...