9-22
P129394-068
Operator Manual
Troubleshooting
Alarm
Description
Screen with Operator Instructions
9.4.9 Door Seal A
Switch Malfunction
Occurs if switched
neutral relay,
associated with door
seal A, closed switch
fails.
Screen with Service Instructions
STATUS . . ALARM!
286
DOOR A SEAL SWITCH MALFUNCTION
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
287
DOOR A SEAL SWITCH MALFUNCTION
-> SEAL SWITCH CLOSED OUT OF CYCLE
CAUSES AND CORRECTIONS:
1. SEAL SWITCH MALFUNCTION
-> CHECK LPS1 (PS2) FOR FAILED CLOSED2.
2. SHORTED CABLE
-> CHECK CABLE AT P19 FOR SHORT CIRCUIT
PAPER
FEED
EXIT
CLEAR
ALARM
Содержание 36 H
Страница 2: ......
Страница 14: ...xii P129394 068 Operator Manual Techniques of Sterilization LIST OF FIGURES Descrption Page ...
Страница 59: ...6 11 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 63: ...6 15 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 66: ...6 18 P129394 068 Operator Manual Sterilizer Operation Figure 6 9 Typical Printout of a Liquid Cycle ...
Страница 67: ...6 19 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 78: ...6 30 P129394 068 Operator Manual Sterilizer Operation Healthcare Capital Equipment ...
Страница 116: ...8 10 P129394 068 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 140: ...9 24 P129394 068 Operator Manual Troubleshooting Healthcare Capital Equipment ...
Страница 152: ...10 12 P129394 068 Operator Manual Service Procedures Healthcare Capital Equipment ...