background image

Users Manual

• Mode d’emploi
• Bedienungshandbuch
• Manual d’Uso
• Manual de uso

TMD-52

Thermocouple 
Thermometer  
K/J/T/E type

Содержание TMD-52

Страница 1: ...Users Manual Mode d emploi Bedienungshandbuch Manual d Uso Manual de uso TMD 52 Thermocouple Thermometer K J T E type ...

Страница 2: ......

Страница 3: ...TMD 52 Thermocouple Thermometer K J T E type Users Manual June 2010 Rev 1 2010 Amprobe Test Tools All rights reserved Printed in Taiwan English ...

Страница 4: ...he exclusion or limitation of an implied warranty or of incidental or consequential damages this limitation of liability may not apply to you Repair All test tools returned for warranty or non warranty repair or for calibration should be accompanied by the following your name company s name address telephone number and proof of purchase Additionally please include a brief description of the proble...

Страница 5: ... ON L4Z 1X9 Tel 877 AMPROBE 267 7623 Tel 905 890 7600 Non Warranty Repairs and Replacement Europe European non warranty units can be replaced by your Amprobe Test Tools distributor for a nominalv charge Please check the Where to Buy section on www amprobe com for a list of distributors near you European Correspondence Address Amprobe Test Tools Europe In den Engematten 14 79286 Glottertal Germany ...

Страница 6: ... J T E Type 14 13 11 12 10 8 7 9 1 3 2 4 5 6 1 Power Button 2 HOLD Button 3 C F Button 4 REL Button 5 Type Button 6 MAX MIN Button 7 Hi Lo Button 8 p Button 9 TC OFFSET Button 10 Button 11 q Button 12 SET Button 13 LCD Display 14 T1 T2 Input ...

Страница 7: ...ature display 2 T2 Temperature display 3 T1 T2 differential 4 Type of T C thermocouple 5 Low BATT display Screen Display 6 MAX MIN reading 7 Data HOLD mode 8 Relative mode 9 SET mode 10 Degrees C F 1 4 6 5 3 2 8 7 9 ...

Страница 8: ... 3 OPERATION INSTRUCTIONS 3 Power Button 3 HOLD Button 3 C F Button 4 REL Button 4 Type Button K J T E Input Thermocouple Type 4 MAX MIN Button Record mode 5 Hi Lo Button LIMITS mode 5 Button 5 TC OFFSET Button 5 Button 6 Button 6 SET Button 6 SPECIFICATION 7 MAINTENANCE 9 ...

Страница 9: ...g voltages at the measurement surface over 24V AC or DC To avoid damage or burns do not make temperature measurement in microwave ovens Repeated sharp flexing can break the thermocouple leads To prolong lead life avoid sharp bends in the leads especially near the connector Unpacking and Inspection Your shipping carton should include 1 TMD 52 meter 1 Manual 2 Thermocouple 4 AAA Batteries 1 Plain wh...

Страница 10: ...or common use Robust protective Holster Auto Power Off and Backlight functions MIN MAX AVG REL HOLD functions OPERATING INSTRUCTIONS Power Button Press the key turns the thermometer on or off In the MAX MIN record mode can not power off must exit MAX MIN record mode then power off In power off status push this key more than 4 seconds to disable auto power off and turn on the power HOLD Button Pres...

Страница 11: ...zero the display and store the displayed reading as a reference value and annunciator REL is displayed Pressing REL key more than 2 seconds to exit the relative mode The relative value can also be entered by the user See SET mode later in this manual When the desired Relative value has been entered press SET key use set Relative value as a reference value Press REL key again to exit the relative m...

Страница 12: ... mode can only be cancelled by pressing and hold down the MAX MIN key for 2 seconds to exit and erased recorded readings Hi Lo Button LIMITS mode Press Hi Lo key to enter the Hi Lo LIMITS comparative mode LIMIT is displayed When the input temperature value more than Hi value the beeper emits a continuity pulse tone and Hi is displayed and when input temperature value less than Lo value The beeper ...

Страница 13: ...blink digit press to change blink digit If have no blink digit press to set the positive or negative for this relative value Then press ENTER key stored the relative value for T1 enter set relative value for T2 In this Relative SET mode the value can not be 1999 9 counts If this value more than 1999 9 counts Err displayed and re input Set Hi Lo Limit value Press SET key to enter SET mode then pres...

Страница 14: ...nk digit If have no blink digit press to set the positive or negative for this TC OFFSET value Then press ENTER key stored the TC OFFSET value for T1 enter set TC OFFSET value for T2 In this TC OFFSET SET mode the value can not be 5 0 counts If this value more than 5 0 counts Err displayed and re input SPECIFICATION Electrical Temperature Scale Celsius or Fahrenheit user selectable Measurement Ran...

Страница 15: ...dg 4 F on 76 F to 364 F Environmental Ambient Operating Ranges 0 C to 50 C 32 F to 122 F 80 R H Storage Temperature 20 C to 60 C 4 F to 140 F 70 R H General Display Automatic positive implied negative polarity indication Over Load OL or OL is displayed Zero Automatic Reading Rate 1 time per second Battery 4 X 1 5V AAA Battery Life 200 hours typical with carbon zinc battery Low battery Indication T...

Страница 16: ... amended by 93 68 EEC CE Marking However electrical noise or intense electromagnetic fields in the vicinity of the equipment may disturb the measurement circuit Measuring instruments will also respond to unwanted signals that may be present within the measurement circuit Users should exercise care and take appropriate precautions to avoid misleading results when making measurements in the presence...

Страница 17: ... remove screw from back of meter and lift off the battery cover 3 Remove the batteries from the battery compartment and replace them with new ones 4 When not use for long time remove the battery 5 Do not store the unit in place with temperature and humidity beyond the recommended storage temperature Cleaning Periodically wipe the case with a damp cloth and detergent do not use abrasives or solvent...

Страница 18: ......