Amprobe TC-253B Скачать руководство пользователя страница 1

User Manual

TC-253B

Temperature Converter

Содержание TC-253B

Страница 1: ...User Manual TC 253B Temperature Converter ...

Страница 2: ... FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES ARISING FROM ANY CAUSE OR THEORY Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages this limitation of liability may not apply to you EXPLANATION OF SYMBOLS Attention Refer to Operating instructions Ground connection Battery WARNING AND PRECA...

Страница 3: ...Temperature Converter from its packing you should have the following items 1 Temperature Converter 2 K type Bead Thermocouple 3 Operator s Manual FRONT PANEL Refer to Figure 1 and the following numbered steps to familiarize yourself with the meter s front panel controls and connectors 1 Red LED Flashing while turned ON 2 Function Switch Slide switch used to select measuring function 3 Output plugs...

Страница 4: ...timeter Set the Multimeter to DC mV or V range 5 Place the sensing point of thermocouple on the surface or environment to be measured 6 Wait until the reading stabilizes GENERAL SPECIFICATIONS This Temperature Converter conforms to the temperature voltage tables of the National Bureau of Standards and to the IEC 584 Standards for K type 1 Temperature Scale Celsius or Fahrenheit user selectable 2 I...

Страница 5: ...ts when making measurements in the presence of electronic interference ENVIRONMENTAL CONDITIONS 1 Indoor Use 2 Operating Ambient 0 to 80 R H 0 to 35 ºC 0 to 70 35 to 50ºC R H 3 Storage Temperature 20 to 60ºC 0 to 80 R H with battery removed from Temperature Converter ELECTRICAL SPECIFICATIONS MEASUREMENT RANGE ACCURACY INPUT PROTECTION 50ºC 20ºC 2 0 rdg 2ºC 19ºC 350ºC 0 5 rdg 2ºC 351º C 500ºC 2 0 ...

Страница 6: ...ther qualified instrument service personnel To maintain a thermocouple in good condition observe the following items Avoid excess bending Don t overheat the thermocouple Avoid chemical reactions that can damage the thermocouple BATTERY CHECKING AND REPLACEMENT To check the battery 1 Turn the TC 253B OFF 2 Insert the output plugs V and COM to the input terminals of multimeter to DCV for autoranging...

Страница 7: ...hould be accompanied by the following information or items company name customer s name address telephone number proof of purchase warranty repairs a brief description of the problem or the service requested and the appropriate service charge for non warranty repairs Please include the test leads with the meter Service charges should be remitted in the form of a check a money order credit card wit...

Страница 8: ...maintenance The information in this document may not be duplicated in any manner without the prior approval in writing from Amprobe Specifications subject to change Amprobe 2007 U S Service Center Amprobe Everett WA 98203 Tel 877 956 2680 Fax 425 446 6390 Canadian Service Center Amprobe 400 Britannia Rd E Unit 1 Mississauga ON L4Z 1X9 Tel 905 890 7600 Fax 905 890 6866 ...