background image

1) “ ” Power
2) °C /°F
3) “ ” Backlight
4) “T1 T2/T1-T2” Main display
5) “TYPE” Main display
6) “MAX/MIN” Main display
7) “REL” Main display
8) “HOLD” Data Hold
9) “SET”

1) Low battery
2) RTD Probe select
3) MAX/MIN/AVG 

record mode

4) HOLD enabled
5) Relative mode

10) “T1 T2 T1-T2” 

Second display

11) “TYPE” Second display
12) “Hi/Lo LIMITS”

Main display

13) Wireless

(RTD-10W Only)

14) Display
15) T1/T2 Input

SCREEN DISPLAY

6) Hi/Lo Limits 

comparative mode

7) Temperature 

measurement units

8) T1/T2 thermocouple 

or T1-T2 differential

9) Wireless function enabled 

(RTD-10W Only)

8

7

9

1

3

2

4

5

6

Содержание RTD-10

Страница 1: ...Users Manual RTD 10 RTD 10W RTD Thermometer RTD Thermometer Wireless For detailed specifications and ordering info go to www TestEquipmentDepot com ...

Страница 2: ...RTD 10 RTD 10W RTD Thermometer RTD Thermometer Wireless Users Manual June 2010 Rev 1 2010 Amprobe Test Tools All rights reserved Printed in Taiwan English ...

Страница 3: ...T INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES ARISING FROM ANY CAUSE OR THEORY Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages this limitation of liability may not apply to you Repair All test tools returned for warranty or non warranty repair or for calibration should be accompanied by the following your na...

Страница 4: ...acement US and Canada Non warranty repairs in the United States and Canada should be sent to a Amprobe Test Tools Service Center Call Amprobe Test Tools or inquire at your point of purchase for current repair and replacement rates ...

Страница 5: ...RTD 10 RTD 10W 15 14 12 10 11 8 7 9 1 3 2 4 5 6 13 Main Display Second Display Third Display ...

Страница 6: ... 3 MAX MIN AVG record mode 4 HOLD enabled 5 Relative mode 10 T1 T2 T1 T2 Second display 11 TYPE Second display 12 Hi Lo LIMITS Main display 13 Wireless RTD 10W Only 14 Display 15 T1 T2 Input Screen Display 6 Hi Lo Limits comparative mode 7 Temperature measurement units 8 T1 T2 thermocouple or T1 T2 differential 9 Wireless function enabled RTD 10W Only 8 7 9 1 3 2 4 5 6 ...

Страница 7: ...y Input Selection 5 TYPE Pt385 Pt3926 Pt3916 Main display Input RTD Probe Select 5 MAX MIN With Time Record Mode 5 REL Relative Mode 6 HOLD Mode 6 SET Mode 7 T1 T2 T1 T2 second display Input Selection 7 TYPE Pt385 Pt3926 Pt3916 Second display Input RTD Probe select 8 Wireless Mode RTD 10W Only 8 RTD PROBE CONNECTION 9 TEMPERATURE VSRESISTANCE TABLE ITS90 10 SPECIFICATION 10 MAINTENANCE 11 ...

Страница 8: ...C Rules Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation Note This equipment RTD 10W has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to p...

Страница 9: ...cted Consult the dealer or an experienced radio TV technician for help Shielded interface cables must be used in order to comply with emission limits Changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment Wireless Note Wireless receiver must keep a distance at least 40cm from the meter and meter to meter dist...

Страница 10: ...or European Curve Alpha 00385 Pt3926 for American Curve Alpha 003926 Pt3916 for Japan Curve Alpha 003916 Features Wireless two ways transmission RTD 10W only Dual input function Pt3916 Pt3926 Pt385 RTD standard MAX MIN with elapsed time AVG REL HOLD function Warning beeper with Hi Lo setting Time setting REL setting According to temperature standard ITS 90 C F user selective Resolution 0 1 C to 0 ...

Страница 11: ... T1 probe T2 probe or the difference between the two probes T1 T2 When the thermometer is turned on it is set to T1 when main display input selected T1 then T1 input can select alternate of probe by pressing Pt385 Pt3926 Pt3916 key switch TYPE Pt385 Pt3926 Pt3916 Main display Input RTD Probe Select The TYPE key switches the T1 input to select the Pt385 Pt3926 or Pt3916 RTD probe as input when main...

Страница 12: ... Reading as a reference value REL is shown on the display Press REL key again to exit the relative mode The relative reference value can also be entered by the user See SET mode later in this manual When the desired relative value has been entered press REL key to enter the relative mode and than press SET key use the entered relative value as a reference value Press REL key again to exit the rela...

Страница 13: ...psed time starts from set time value 3 Hi Limit value set mode Press ENTER key to skip Hi Limit value set mode is displayed in main display Hi Limit value is entered via overlay numbers then press ENTER key to store the Hi Limit value is displayed in main display Lo Limit value is entered via overlay numbers then press overlay ENTER key to store the Lo Limit value and exit SET mode 4 When the ther...

Страница 14: ...ne Press Hi Lo LIMITS key again to exit the Hi Lo Limits mode Wireless Mode RTD 10W Only Press the key for more than two seconds to start wireless function Press the key again for another two seconds to stop wireless function The wireless mode will shut down if there is no wireless signal for two minutes To SET CH ID to 00 00 press the Hi Lo LIMITS key and power key for more than 6 seconds with th...

Страница 15: ...O 100 000O 100 C 138 505O 139 272O 139 171O 200 C 175 856O 177 362O 177 155O 300 C 212 052O 214 275O 213 957O 400 C 247 092O 250 018O 249 584O 500 C 280 977O 284 591O 284 036O 600 C 313 708O 317 994O 317 313O 700 C 345 280O 800 C 375 700O MINI DIN PLUG 3 Wires 4 Wires Front View Rear View RTD 100 Pt385 Pt3916 Pt3926 RTD 100 Pt385 Pt3916 Pt3926 G 2 1 2 1 2 1 2 1 ...

Страница 16: ... rdg 0 2 C on C scale 0 05 rdg 0 4 F on F scale Temperature Coefficient 0 1 times the applicable accuracy specification per C from 0 C to 18 C and 28 C to 50 C 32 F to 64 F and 82 F to 122 F Input Protection 24V dc or 24V ac rms maximum input Voltage on any combination of input pins Input Connector Accepts for 3 wires or 4 wires Mini Din plug and the meter is designed for standard The other probes...

Страница 17: ...ize Battery Life 200 hours typical with carbon zinc battery Dimensions 160mm H x 83mm W x 38mm D 6 3 in H x 3 3 in W x 1 5 in D Weight Approx 260g 0 6lb including batteries Wireless Features Frequency range 910 920MHz RTD 10W Only Low current consumption less than 1 mA The transmitting distance can reach 25 M without magnetic interference EMC Conforms to EN61326 1 This product complies with requir...

Страница 18: ...ent within the measurement circuit Users should exercise care and take appropriate precautions to avoid misleading results when making measurements in the presence of electronic interference MAINTENANCE AND REPAIR Warning To avoid possible electrical shock disconnect the thermocouple connectors from the thermometer before removing the cover Installing and Replacing Battery A Screw B Battery Cover ...

Страница 19: ...nd lift off the battery cover 3 Remove the batteries from the battery compartment and replace them with new ones 4 When not use for long time remove the battery 5 Do not store the unit in place with temperature and humidity beyond the recommended storage temperature Cleaning Periodically wipe the case with a damp cloth and detergent do not use abrasives or solvents ...