ROUTINE MAINTENANCE
26
D-0014930-A
– Otowave 102-C Operating Manual
13.
ROUTINE MAINTENANCE
13.1.
CLEANING THE OTOWAVE
The Otowave is a precision instrument. Handle it carefully in order to ensure its continued accuracy and
service. Before cleaning switch off the instrument. Use a soft damp cloth and mild detergent to clean the
instrument panel and case. Ensure no moisture enters the instrument.
13.2.
EAR TIP AND PROBE
Ear tips should be replaced after a single use.
The probe tip and its associated seal are disposable devices.
The probe tip should be checked before each ear insertion to ensure it is undamaged and that none of the
tubes through it are blocked. It should be replaced if necessary.
The seal should be replaced when the probe tip is replaced, if it shows signs of wear, or if a pressure leak is
suspected.
Handle the probe and accessories with care. Do not allow moisture,
condensation, fluids or debris to enter the probe.
13.3.
CALIBRATION AND REPAIR OF THE INSTRUMENT
Amplivox recommends that the Otowave 102-C is calibrated annually. Please contact Amplivox for details.
If the instrument is to be used at elevations above that specified re-calibration must be undertaken at the
intended operating elevation.
The instrument should be returned to the manufacturer for service & repair.
There are no user-serviceable parts within it.
Please use the original shipping carton and packaging to transport the instrument. Place the instrument in a
plastic bag before packing to prevent dirt and dust getting into the probe.
WARNING
WARNING