TAKING MEASUREMENTS

 

 

 

 

                                                             14 

D-0014930-A 

– Otowave 102-C Operating Manual 

                                

 

    

REFLEX SUMMARY    

dB 
100 

    

    

X

    - 

90  

    

X

    

    

 

80  

X

    

    

    

 

70  

X

    

    

X

    

X

 

 
Hz 500   1k   2k   4k

            

Press  ◄  to  return  and  view  the  tympanogram,  reflex  results  or  to  repeat  the  test.  When  satisfied  with  the 
results press ►.  

The message “Saving as last test” will be displayed and the results will be saved in the “last test” memory. The 
results will remain available until a new test is started, even if the Otowave is turned off. 
 
If both ears were chosen for test the entire sequence will now be repeated for the right ear:  
 

TESTING RIGHT EAR     
 
INSERT PROBE    
 
 
 
Cancel            Skip            

Press ► to skip testing of the right ear and display the PROCESS RESULTS menu. Press ◄ to cancel and 
return to the ear selection menu. In both cases the left ear results are retained and may be viewed as the LAST 
TEST. 

Otherwise insert the probe; the right ear test will then proceed as described above. 
 
When  the  selected  ears  have  been  tested  and  the  results  saved  the  PROCESS  RESULTS  menu  will  be 
displayed. This accesses the following functions: 

 

Print the results 

 

Save the results in the internal database 

 

Review the results as described above 

 

Return to the main menu 

The results of the last test performed remain available even if the Otowave has been turned off. To view these 
results select VIEW THE LAST TEST from the main menu. After selecting the required ear the tympanogram 
will be displayed. It will then be possible to view the results and select the PROCESS RESULTS menu as if 
the test had just been completed. 

 Please note:

 Results of the last test will be erased as soon as a new test is started. Test results should 

be saved to the Otowave’s database, printed or sent to a computer to ensure that data is not lost. 

5.4.

 

EAR SEAL CHECK 

The type of ear seal check employed at the start of a test can be set in the CONFIGURATION menu (Section 
6).  The  default  STANDARD  option  is  adequate  for  most  tests,  although  it  may  not  always  be  possible  to 
generate the extremes of pressure with this setting. 
 
However, if difficulty is experienced in using the ear tips to create a seal the alternative  EXTENDED option 
may be helpful. This checks that a range of pressures will be available before starting a test by means of a 
visual indication of the quality of the seal: 

 

Содержание Otowave 102-C

Страница 1: ...Otowave 102 C Operating Manual...

Страница 2: ...ivox Ltd 3800 Parkside Solihull Parkway Birmingham Business Park Birmingham West Midlands B37 7YG United Kingdom www amplivox com For all enquiries please contact us Amplivox Ltd 10393 West 70th Stree...

Страница 3: ...If supplied in the US please note Caution Federal Law restricts this device to sale by or on the order of a licenced medical professional...

Страница 4: ...e 6 4 1 Battery PACK 6 4 2 Operating language 6 4 3 The Cradle 6 4 3 1 General 6 4 3 2 Connectors 6 4 3 3 Indicators 7 4 4 Controls and indicators Handheld Unit 8 4 5 The probe 9 4 6 Start up and menu...

Страница 5: ...Maintenance 26 13 1 Cleaning the Otowave 26 13 2 Ear tip and Probe 26 13 3 Calibration and Repair of the Instrument 26 14 Error Messages Fault Conditions 27 15 Technical Specification 29 15 1 Perform...

Страница 6: ...in non volatile memory Configurable settings for user preferences held in non volatile memory Printout of data to a printer Data transfer to a computer via USB for storage viewing printing using eith...

Страница 7: ...ip Otowave 14mm 25 pieces 8012979 Ear tip Otowave 15mm 25 pieces 8012981 Ear tip Otowave 19mm 25 pieces 8012983 1 6 GUARANTEE All Amplivox instruments are guaranteed against faulty materials and manuf...

Страница 8: ...trument and its accessories and the function of single use parts Do not use the instrument in an oxygen rich environment or in the presence of a flammable anaesthetic mixture or other flammable agents...

Страница 9: ...ng a microphone and the admittance calculated from the result In line with normal audiometric practice admittance is displayed as an equivalent volume of air in ml for 226 Hz The residual ear canal vo...

Страница 10: ...e to become stiffer or a preset maximum level is reached When the change in admittance exceeds a predetermined threshold this constitutes a reflex and the change in admittance at that level when the s...

Страница 11: ...onfiguration the contents of the database the calibration settings or the results of the last test 4 2 OPERATING LANGUAGE To set the operating language English French Spanish Italian Portugese or Germ...

Страница 12: ...ter USB USB connector Type B Computer via USB port For connected parts marked make sure to only connect the parts or accessories supplied with the instrument or supplied by Amplivox or an Amplivox dis...

Страница 13: ...r go to the next step Press the left navigation key to cancel an operation or go back to the previous step The function of the left and right keys is usually shown on the bottom line of the display Wh...

Страница 14: ...s blocked or damaged Do not remove the nut securing the boss to the body of the instrument Please note When replacing the probe tip ensure that the seal is correctly located with the flat side aligned...

Страница 15: ...ting Manual 4 7 INITIAL SETTINGS Use the CONFIGURATION options see Section 6 to select the following options as required display contrast for ease of viewing correct local date and time date format DD...

Страница 16: ...se note The probe tip must be fitted with a new ear tip before it is presented to a patient s ear canal The ear tip must be fitted completely to the probe tip and must not occlude any of the four hole...

Страница 17: ...default this test is only performed if a peak is found in the tympanogram This and other reflex test options may be changed in the CONFIGURATION menu see Section 6 Before starting the reflex test the...

Страница 18: ...elect an alternative peak using the and keys The figures displayed will change to reflect the peak selected and will be saved with the tympanogram To repeat the test press When satisfied with the tymp...

Страница 19: ...ns Print the results Save the results in the internal database Review the results as described above Return to the main menu The results of the last test performed remain available even if the Otowave...

Страница 20: ...ISPLAYED INDICATOR STATUS LIKELY CAUSE S WITHDRAW PROBE Yellow Flashing The probe has been moved during measurement Re insert the probe to repeat the test Volume outside range WITHDRAW PROBE Yellow Fl...

Страница 21: ...eload Defaults Reset the sweep settings to its original settings 6 2 REFLEX OPTIONS Amplivox YouTube Video available on how to add ipsilateral reflexes to test protocol ITEM DESCRIPTION DEFAULT Level...

Страница 22: ...t produces a response By setting REFLEX AUTO STOP to NO the Otowave 102 C will test for a reflex at all selected levels Note that 100dBHL at 4000Hz is not available No Polarity Define the polarity of...

Страница 23: ...epartment Allows the department name to be entered The name will appear at the top of the print out Defaults Reset the system settings to the original settings Language Change the operation language t...

Страница 24: ...stored records this same identifier may be used for different tests for the same patient 7 1 DATA ENTRY PATIENT INITIALS _____ ABCDEFGHIJKLM NOPQRSTUVWXYZ 01233456789 Hold to enter cancel To enter the...

Страница 25: ...ults to a dedicated thermal printer or a suitably equipped computer via a USB link When printing directly to the printer the data is received through a cable connected between the cradle and the print...

Страница 26: ...the analysis and the results The name of the hospital the department and the calibration dates for the instrument may also be printed if required There is space for additional details to be handwritte...

Страница 27: ...st select LIST RECORDS All other options operate on groups of records 10 1 LIST RECORDS LIST RECORDS shows the stored tests 6 at a time most recent first Records Stored 6 30 ABC 02 01 06 14 15 2 DEF 3...

Страница 28: ...DELETE RECORDS allows a group of records to be deleted It is possible to delete all records all records that have been printed or all records that have been sent to a computer Confirmation of the dele...

Страница 29: ...nto the hole at the 2ml end of the test cavity Make sure that the probe is pushed in fully and is held tight against the stop The probe must be square to the end of the test cavity The display should...

Страница 30: ...C Operating Manual 12 SYSTEM INFORMATION 1 Battery Remaining battery capacity 2 Last Cal Last calibration date 3 Next Cal Next calibration date 4 Serial No Serial number of Otowave 5 Ver Firmware vers...

Страница 31: ...d be replaced if necessary The seal should be replaced when the probe tip is replaced if it shows signs of wear or if a pressure leak is suspected Handle the probe and accessories with care Do not all...

Страница 32: ...ast measured data cannot be found under VIEW THE LAST TEST NEW TEST might have been selected in between and thereby deleted the last test s from the short term memory Data you wish to be stored should...

Страница 33: ...IRED Indicator LED c steady light The current date is later than the next calibration date Check that the clock is set to the correct date If so arrange for the instrument to be recalibrated Tests can...

Страница 34: ...95dB 90dB or 85dB with 5 dB steps Reflex analysis Reflex pass fail at each level tested maximum amplitude of each reflex nominal pressure used for the reflex test computer display only Pressure used...

Страница 35: ...rent 70mA Current while testing 230mA Physical Display 128 x 64 pixels 8 lines of 21 characters Dimensions 230mm L x 115mm W x 70mm H Total Weight handset and cradle 650g Environmental Operating tempe...

Страница 36: ...dby condition Definition Refer to instruction manual mandatory Definition Type BF applied part an applied part providing a higher degree of protection against electric shock than that provided by a Ty...

Страница 37: ...ion of waste electrical items from the general waste stream and into reuse recovery and recycling routes Therefore for any waste electrical units purchased from Amplivox that either bear the crossed o...

Страница 38: ...tly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC 61000 3 2 Not applicable Voltage fluctuations flicker emissions I...

Страница 39: ...ines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Not applicable Not applicable Power frequency 50 60 Hz...

Страница 40: ...F transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the...

Страница 41: ...o the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 8...

Страница 42: ...s to ensure that the general requirements of clause 16 of IEC 60601 1 2005 are met External equipment intended for connection to signal input signal output or other connectors shall comply with the re...

Страница 43: ...ELECTRICAL EQUIPMENT 38 D 0014930 A Otowave 102 C Operating Manual Diagram 1 Otowave 102 C used with the designated printer Mains Outlet Printer via USB Cradle Connection Printer Power Supply Otowave...

Страница 44: ...LECTRICAL EQUIPMENT 39 D 0014930 A Otowave 102 C Operating Manual Diagram 2 Otowave 102 C used with a computer USB connection Mains Outlet Computer via USB Cradle Connection Computer Power Supply Otow...

Страница 45: ...NT 40 D 0014930 A Otowave 102 C Operating Manual Copyright 2020 Amplivox Ltd All rights reserved No part of this publication may be reproduced or transmitted in any form or by any means without the pr...

Отзывы: