USING THE OTOWAVE
9
D-0014930-A
– Otowave 102-C Operating Manual
4.5.
THE PROBE
1
Boss and Nut
Connection on probe body for attaching nose cone
2
Seal gasket
Gasket used to ensure air flow
3
Probe tip
Transparent probe tip which houses the gasket
4
Nose cone
Top part of probe to securely fasten probe tip and gasket
The small holes through the Otowave probe tip must be kept clear. If these become blocked a warning
message will be displayed. The probe tip must be removed and replaced.
To remove the probe tip, unscrew the nose cone and remove the probe tip from the boss. A small seal will be
found in the base of the probe tip. This should be examined and replaced if it is blocked or damaged. Do not
remove the nut securing the boss to the body of the instrument.
Please note:
When replacing the probe tip, ensure that the seal is correctly located with the flat side
aligned with the flat side within the base of the probe tip. Push the probe tip over the boss and replace the
nose cone. Make sure that the nose cone is screwed home firmly but do not over-tighten. Do not use any tools
to tighten the nose cone.
After replacing the tip a Daily Check should be carried out (see Section 11).
4.6.
START-UP AND MENU DISPLAYS
When the Otowave 102-C is turned on the start-up screen is shown while internal tests are performed and the
pump is initialised. When the start-up sequence is complete the MAIN MENU is displayed:
MAIN MENU
NEW TEST
CONFIGURATION
VIEW THE LAST TEST
Select
Use the navigation keys to scroll through and select menu options.